Temporal evolution of low-coherence reflectrometry signals in photoacoustic remote sensing microscopy

Appl Opt. 2017 Jun 20;56(18):5172-5181. doi: 10.1364/AO.56.005172.

Abstract

Recently, a new noncontact reflection-mode imaging modality called photoacoustic remote sensing (PARS) microscopy was introduced providing optical absorption contrast. Unlike previous modalities, which rely on interferometric detection of a probe beam to measure surface oscillations, the PARS technique detects photoacoustic initial pressures induced by a pulsed laser at their origin by monitoring intensity modulations of a reflected probe beam. In this paper, a model describing the temporal evolution from a finite excitation pulse is developed with consideration given to the coherence length of the interrogation beam. Analytical models are compared with approximations, finite-difference time-domain (FDTD) simulations, and experiments with good agreement.