Big Chemical Encyclopedia

Chemical substances, components, reactions, process design ...

Articles Figures Tables About

Wavelength reproducibility

Figure 15-23. Calculalcil value ol ihc square of (lie normalized electrical field i 2 at the Cw/PEOPT interface lor PEOPT thickness of 30 nm (solid line) and 40 nm (dashed line) versus thickness of the CM layer at a wavelength of 460 nin. The inset shows the calculated distribution of the square of Ihc normalized opiical field E ] inside an 1TO (120 mn)/PEDOT-PSS (110 iini)/PEOPT(30 nm)/CW) (32 nm) device at the same wavelength (reproduced by permission of Wilcy-VCH from Ref. 92 ). Figure 15-23. Calculalcil value ol ihc square of (lie normalized electrical field i 2 at the Cw/PEOPT interface lor PEOPT thickness of 30 nm (solid line) and 40 nm (dashed line) versus thickness of the CM layer at a wavelength of 460 nin. The inset shows the calculated distribution of the square of Ihc normalized opiical field E ] inside an 1TO (120 mn)/PEDOT-PSS (110 iini)/PEOPT(30 nm)/CW) (32 nm) device at the same wavelength (reproduced by permission of Wilcy-VCH from Ref. 92 ).
Figure 3.3 Near-field transmission spectra and images of a single gold nanorod (length 510nm, diameter 20nm). The two transmission spectra were obtained at positions 1 and 2 indicated in the inset. Each image was obtained at the resonance peak wavelength. (Reproduced with permission from Royal Society of Chemist [10]). Figure 3.3 Near-field transmission spectra and images of a single gold nanorod (length 510nm, diameter 20nm). The two transmission spectra were obtained at positions 1 and 2 indicated in the inset. Each image was obtained at the resonance peak wavelength. (Reproduced with permission from Royal Society of Chemist [10]).
Figure 6.4 Plotting absorbance versus wavelength at several levels of pH for a BCG optical sensor shows that the alkaline form of the indicator absorbs at about 620 nm and the acidic form absorbs at about 450 nm. At about 550 nm both forms have the same molar absorptivity (isobestic wavelength). (Reproduced from LASERFOCUSWORLD.COM.)... Figure 6.4 Plotting absorbance versus wavelength at several levels of pH for a BCG optical sensor shows that the alkaline form of the indicator absorbs at about 620 nm and the acidic form absorbs at about 450 nm. At about 550 nm both forms have the same molar absorptivity (isobestic wavelength). (Reproduced from LASERFOCUSWORLD.COM.)...
The temperature tuning coefficient of TeOa is 0.025 nm/°C, so it is clear that in order to combat the effects of thermal heating of the crystal due to incident optical and aconstic power, as well as environmental considerations, a good level of thermal control is needed. This can be in the form of a thermo-electrically cooled enclosure. This will deal to a certain extent with the need to achieve analyzer wavelength repeatability. However, it does not address the issue of wavelength reproducibility between AOTF modnles, which impacts the transportability of calibrations and datasets between analyzers. [Pg.126]

Fig. 17 SEM images of structures fabricated in a triacrylate resin at threshold powers using a 730 nm excitation of r.l and b 520 nm excitation of r.3. The lower panels show the dimensions, transversal to the beam propagation direction, of single lines of the strnc-tures for each excitation wavelength. Reproduced with permission from [179], 2007, Optical Society of America... Fig. 17 SEM images of structures fabricated in a triacrylate resin at threshold powers using a 730 nm excitation of r.l and b 520 nm excitation of r.3. The lower panels show the dimensions, transversal to the beam propagation direction, of single lines of the strnc-tures for each excitation wavelength. Reproduced with permission from [179], 2007, Optical Society of America...
Wavelength reproducibility Stray light Potassium chloride solution (200 nm) Potassium iodide solution V V V V... [Pg.154]

Figure 11-6. Lifetimes of 1-methyl uracil, 1,3-dimethyl uracil, 1,3-dimethyl thymine, and thymine at different excitation wavelengths. (Reproduced with permission from J. Phys. Chem. 2004, 108, 943-949. Copyright 2004 American Chemical Society.)... Figure 11-6. Lifetimes of 1-methyl uracil, 1,3-dimethyl uracil, 1,3-dimethyl thymine, and thymine at different excitation wavelengths. (Reproduced with permission from J. Phys. Chem. 2004, 108, 943-949. Copyright 2004 American Chemical Society.)...
Figure 12 Excitation spectrum of the liuninescence of T1+ in KF (variation of the emission intensity as a function of the excitation wavelength). (Reproduced by permission of Ref 67)... Figure 12 Excitation spectrum of the liuninescence of T1+ in KF (variation of the emission intensity as a function of the excitation wavelength). (Reproduced by permission of Ref 67)...
Figure 3.48. UV transmission characteristics of dyed PMMA and Kodak 809 resist as a function of wavelength. Reproduced with permission from r erence 173. Copyright 1984 Society of Photo-Optical Instrumentation Engineers.)... Figure 3.48. UV transmission characteristics of dyed PMMA and Kodak 809 resist as a function of wavelength. Reproduced with permission from r erence 173. Copyright 1984 Society of Photo-Optical Instrumentation Engineers.)...
Fig. 7.23 In-line transmittance spectra of the polycrystalline MgAl204 sintered by SPS at different preload pressures PI. The theoretical transmittance of a single-crystal MgAl204 was calculated based on the refraction index as a function of wavelength. Reproduced with permission from [140]. Copyright 2009, Elsevier... Fig. 7.23 In-line transmittance spectra of the polycrystalline MgAl204 sintered by SPS at different preload pressures PI. The theoretical transmittance of a single-crystal MgAl204 was calculated based on the refraction index as a function of wavelength. Reproduced with permission from [140]. Copyright 2009, Elsevier...
Wavelength reproducibility can be simply checked by an overlay of repetitively recorded line emitter or holmium perchlorate solution spectra. Since wavelength reproducibility is especially important in kinetic investigations, a chemical reaction can also be used as standard, for which the... [Pg.440]

Figure 24. Sulfone reaction spectrum for characterization of wavelength reproducibility of a spectrometer... Figure 24. Sulfone reaction spectrum for characterization of wavelength reproducibility of a spectrometer...
Spectrofluorimeter. Preferably capable of synchronous scanning, with wavelength reproducibility of excitation and emission monochromators +2nm or better typical monochromator gratings of 6001ines/nm blazed at 300 nm or better, spectral resolution with bandpass of 2.5 nm or less, maximum bandpass of 10 nm and a response of S-20 or S-5 on the photomultiplier tube. [Pg.469]

A second advantage of Fourier transform instruments is their extremely high resolving power and wavelength reproducibility that make possible the analysis of complex spectra in which the sheer number of lines and spectral overlap make the determination of individual spectral features difficult. Figure 7-40. which is papt of an emission spectrum for a steel, illustrates this advantage. The spectrum, which extends from only 299.85 to 300.75 nm, contains thirteen well-separated lines of three elements. The wavelength resolution (AA/A) for the closest pair of lines is about 6 ppm. [Pg.638]

Figure 3.2 Far-infrared absorption spectrum of HCI gas. The vertical coordinate is light intensity transmitted by the gas sample the horizontal coordinate is wavelength. Reproduced by permission from D. Bloor et al., Proc. London Ser. A 260 510 (1961). Figure 3.2 Far-infrared absorption spectrum of HCI gas. The vertical coordinate is light intensity transmitted by the gas sample the horizontal coordinate is wavelength. Reproduced by permission from D. Bloor et al., Proc. London Ser. A 260 510 (1961).
Fig. 2.20. Photoionization schemes, (a) Resonant 1 -i-1 REMPI, (b) nonresonant MPI, resonant 1 + 2 REMPI, and resonant 1 + 1 REMPI with two different wavelengths. Reproduced from Ref. [92] with permission. The Vacuum Society of Japan, 2007. Fig. 2.20. Photoionization schemes, (a) Resonant 1 -i-1 REMPI, (b) nonresonant MPI, resonant 1 + 2 REMPI, and resonant 1 + 1 REMPI with two different wavelengths. Reproduced from Ref. [92] with permission. The Vacuum Society of Japan, 2007.
Figure 4. Effect of silver particle s shape on the peak plasmon resonance scattering wavelength. Reproduced from [14] with permission from the publisher. Figure 4. Effect of silver particle s shape on the peak plasmon resonance scattering wavelength. Reproduced from [14] with permission from the publisher.
Ffe. 7.5 Raman spectra of whole blood samples, a NRS spectinm of dried whole blood, b SERS spectrum of fresh whole blood, c SERS spectrum of fresh plasma. 785 nm excitation wavelength (reproduced with permission from Premasiri et al. 2012a. Copyright 2012 American Chemical Society)... [Pg.160]


See other pages where Wavelength reproducibility is mentioned: [Pg.305]    [Pg.383]    [Pg.36]    [Pg.196]    [Pg.521]    [Pg.215]    [Pg.6074]    [Pg.123]    [Pg.6073]    [Pg.474]    [Pg.521]    [Pg.110]    [Pg.237]    [Pg.619]    [Pg.73]    [Pg.78]    [Pg.4469]    [Pg.4470]    [Pg.107]    [Pg.15]    [Pg.1990]    [Pg.245]    [Pg.72]    [Pg.376]    [Pg.220]    [Pg.37]    [Pg.561]   
See also in sourсe #XX -- [ Pg.205 , Pg.474 ]




SEARCH



Reproducibility

Reproducible

© 2024 chempedia.info