Accuracy and acceptability of the 6-day Enlite continuous subcutaneous glucose sensor

Diabetes Technol Ther. 2014 May;16(5):277-83. doi: 10.1089/dia.2013.0222. Epub 2014 Apr 7.

Abstract

Objective: This study evaluated the performance and acceptability of the Enlite(®) glucose sensor (Medtronic MiniMed, Inc., Northridge, CA).

Subjects and methods: Ninety adults with type 1 or type 2 diabetes wore two Enlite sensors on the abdomen and/or buttock for 6 days and calibrated them at different frequencies. On Days 1, 3, and 6, accuracy was evaluated by comparison of sensor glucose values with frequently sampled plasma glucose values collected over a 12-h period. Accuracy was assessed at different reference glucose concentrations and during times when absolute glucose concentration rates of change were <1, 1-2, and >2 mg/dL/min. The sensor's ability to detect hypoglycemia or hyperglycemia was evaluated with simulated alerts. Subject satisfaction was evaluated with a 7-point Likert-type questionnaire, with a score of 7 indicating strong agreement.

Results: With abdomen sensors under actual-use calibration (mean, 2.8 ± 0.9 times/day), the overall mean (median) absolute relative difference (ARD) values between sensor and reference values were 13.6% (10.1%); the corresponding buttock sensor ARD values were 15.5% (10.5%). With abdomen sensors under minimal calibration (mean, 1.2 ± 0.9 times/day), the mean (median) ARD values were 14.7% (10.8%). Mean ARD values of abdomen sensors at rates of change of <1, 1-2, and >2 mg/dL/min were 13.6%, 12.9%, and 16.3%, respectively. With abdomen sensors, 79.5% and 94.1% of hypoglycemic and hyperglycemic events, respectively, were correctly detected; 81.9% and 94.9% of hypoglycemic and hyperglycemic alerts, respectively, were confirmed. The failure rates for abdomen and buttock sensors were 19.7% and 13.9%, respectively. Mean responses to survey questions for all subjects related to comfort and ease of use were favorable.

Conclusions: The Enlite sensor provided accurate data at different glucose concentrations and rates of change. Subjects found the sensor comfortable and easy to use.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / standards
  • Diabetes Mellitus, Type 1 / blood*
  • Diabetes Mellitus, Type 2 / blood*
  • Female
  • Humans
  • Male
  • Patient Satisfaction
  • Reproducibility of Results
  • Surveys and Questionnaires
  • Young Adult

Substances

  • Blood Glucose