Can STOP-Bang and Pulse Oximetry Detect and Exclude Obstructive Sleep Apnea?

Anesth Analg. 2018 Sep;127(3):736-743. doi: 10.1213/ANE.0000000000003607.

Abstract

Background: Obstructive sleep apnea (OSA) is related to postoperative complications and is a common disorder. Most patients with sleep apnea are, however, undiagnosed, and there is a need for simple screening tools. We aimed to investigate whether STOP-Bang and oxygen desaturation index can identify subjects with OSA.

Methods: In this prospective, observational multicenter trial, 449 adult patients referred to a sleep clinic for evaluation of OSA were investigated with ambulatory polygraphy, including pulse oximetry and the STOP-Bang questionnaire in 4 Swedish centers. The STOP-Bang score is the sum of 8 positive answers to Snoring, Tiredness, Observed apnea, high blood Pressure, Body mass index >35 kg/m, Age >50 years, Neck circumference >40 cm, and male Gender.

Results: The optimal STOP-Bang cutoff score was 6 for moderate and severe sleep apnea, defined as apnea-hypopnea index (AHI) ≥15, and the sensitivity and specificity for this score were 63% (95% CI, 0.55-0.70) and 69% (95% CI, 0.64-0.75), respectively. A STOP-Bang score of <2 had a probability of 95% (95% CI, 0.92-0.98) to exclude an AHI >15 and a STOP-Bang score of ≥6 had a specificity of 91% (95% CI, 0.87-0.94) for an AHI >15. The items contributing most to the STOP-Bang were the Bang items. There was a positive correlation between AHI versus STOP-Bang and between AHI versus oxygen desaturation index, Spearman ρ 0.50 (95% CI, 0.43-0.58) and 0.96 (95% CI, 0.94-0.97), respectively.

Conclusions: STOP-Bang and pulse oximetry can be used to screen for sleep apnea. A STOP-Bang score of <2 almost excludes moderate and severe OSA, whereas nearly all the patients with a STOP-Bang score ≥6 have OSA. We suggest the addition of nightly pulse oximetry in patients with a STOP-Bang score of 2-5 when there is a need for screening for sleep apnea (ie, before surgery).

Trial registration: ClinicalTrials.gov NCT02293421.

Publication types

  • Multicenter Study
  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Body Mass Index
  • Female
  • Humans
  • Hypertension / diagnosis*
  • Hypertension / epidemiology
  • Hypertension / physiopathology
  • Male
  • Middle Aged
  • Oximetry / methods
  • Oximetry / standards*
  • Polysomnography / methods
  • Polysomnography / standards
  • Preoperative Care / methods
  • Preoperative Care / standards*
  • Prospective Studies
  • Sleep Apnea, Obstructive / diagnosis*
  • Sleep Apnea, Obstructive / epidemiology
  • Sleep Apnea, Obstructive / physiopathology
  • Snoring / diagnosis*
  • Snoring / epidemiology
  • Snoring / physiopathology
  • Surveys and Questionnaires / standards*
  • Sweden / epidemiology

Associated data