A comparison of techniques to optimize measurement of voltage changes in electrical impedance tomography by minimizing phase shift errors

AJ Fitzgerald, DS Holder, L Eadie… - IEEE transactions on …, 2002 - ieeexplore.ieee.org
AJ Fitzgerald, DS Holder, L Eadie, C Hare, RH Bayford
IEEE transactions on medical imaging, 2002ieeexplore.ieee.org
In electrical impedance tomography, errors due to stray capacitance may be reduced by
optimization of the reference phase of the demodulator. Two possible methods,
maximization of the demodulator output and minimization of reciprocity error have been
assessed, applied to each electrode combination individually, or to all combinations as a
whole. Using an EIT system with a single impedance measuring circuit and multiplexer to
address the 16 electrodes, the methods were tested on resistor-capacitor networks, saline …
In electrical impedance tomography, errors due to stray capacitance may be reduced by optimization of the reference phase of the demodulator. Two possible methods, maximization of the demodulator output and minimization of reciprocity error have been assessed, applied to each electrode combination individually, or to all combinations as a whole. Using an EIT system with a single impedance measuring circuit and multiplexer to address the 16 electrodes, the methods were tested on resistor-capacitor networks, saline-filled tanks and humans during variation of the saline concentration of a constant fluid volume in the stomach. Optimization of each channel individually gave less error, particularly on humans, and maximization of the output of the demodulator was more robust. This method is, therefore, recommended to optimize systems and reduce systematic errors with similar EIT systems.
ieeexplore.ieee.org
Showing the best result for this search. See all results