Paper
15 March 2019 Comparison of benchtop pressure gradient measurements in 3D printed patient specific cardiac phantoms with CT-FFR and computational fluid dynamic simulations
Author Affiliations +
Abstract
Purpose: Various CT-FFR methods are being proposed as a non-invasive method to estimate cardiac disease severity. 3D printed patient specific cardiovascular models with high geometric accuracy can be used to simulate blood flow conditions and perform precise and repeatable benchtop flow experiments for validation of such methods. Materials and Methods: Twelve patient-specific 3D printed cardiac phantoms were created from CT Angiography (CTA) scans using a compliant 3D printing material. Pressure sensors were connected to the aortic root and distal ends of the three main coronary arteries to measure benchtop pressure gradients for each stenosed vessel. The patient geometries were used in Canon Medical Systems 1D fluid dynamics algorithm to calculate the CT- FFR. In addition, a 3D computational fluid dynamics simulation was done using ANSYS to estimate pressure gradients across the coronary arteries. Experimental data and 1D and 3D flow simulations were compared to the standard catheter lab FFR measurement (Invasive-FFR). Results: The average percent difference in Benchtop FFR/Invasive FFR, CT-FFR/Invasive FFR, and CFDFFR/Invasive FFR was 0.05, 0.06, and 0.07 respectively. The average time it took for the CT-FFR simulation was ~35 minutes and it took ~15 hours for the CFD-FFR simulation but can vary based on the number of iterations the user defines the software to run. Conclusions: Benchtop FFR proved to be highly accurate when compared to both 1D and 3D CFD software and therefore, 3D printing of patient specific coronary phantoms is a quality tool for CT-FFR software validation.
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Kelsey N. Sommer, Lauren Shepard, Vijay Iyer M.D., Erin Angel, Michael F. Wilson, Frank J. Rybicki, Dimitrios Mitsouras, Kanako Kunishima Kumamaru, Stephen Rudin, and Ciprian N. Ionita "Comparison of benchtop pressure gradient measurements in 3D printed patient specific cardiac phantoms with CT-FFR and computational fluid dynamic simulations", Proc. SPIE 10953, Medical Imaging 2019: Biomedical Applications in Molecular, Structural, and Functional Imaging, 109531P (15 March 2019); https://doi.org/10.1117/12.2512118
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
3D modeling

Arteries

3D metrology

Computational fluid dynamics

Computed tomography

Gold

Computer simulations

Back to Top