This book is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) ( http://creativecommons.org/licenses/by-nc-nd/4.0/ ), which permits others to distribute the work, provided that the article is not altered or used commercially. You are not required to obtain permission to distribute this article, provided that you credit the author and journal.
NCBI Bookshelf. A service of the National Library of Medicine, National Institutes of Health.
StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing; 2024 Jan-.
StatPearls [Internet].
Show detailsContinuing Education Activity
Ophthalmometer or keratometer is an instrument used to measure the curvature of the anterior corneal surface and assess the extent and axis of astigmatism. Keratometry is the measurement of the curvature of the anterior corneal surface, usually across a fixed chord length of 2 to 3 mm, which falls within the spherical optic zone of the cornea. The principle of keratometry is that the anterior corneal surface behaves as a convex mirror and the size of the image change with the curvature. There are various clinical uses of keratometers, such as corneal astigmatism measurement, contact lens fitting, assessment of refractive error, IOL power calculation, etc. The keratometer readings may not always be perfect because of mistakes in keratometry. The activity deals with keratometers, keratometry principle, type of keratometers, clinical use, errors in keratometry, and clinical significance of keratometers.
Objectives:
- Describe how to perform keratometry.
- Review the interfering factors for a perfect keratometry.
- Summarize the indications of keratometry by using a keratometer.
- Explain the clinical significance of a keratometer.
Introduction
Keratometer or ophthalmometer is an instrument used to measure the anterior corneal curvature and the axis and extent of astigmatism.[1] The process of these measurements is called keratometry. Hermann von Helmholtz invented the first keratometer, and it came into vogue in 1851.[2] The earlier keratometer was designed by Jesse Ramsden and Everard Home in 1796. In keratometry, the measurements are performed at a fixed chord length of 2 to 3 mm, which falls within the spherical zone of the cornea.[3]
The keratometry is based on the principle that the anterior corneal surface behaves like a mirror and the image size varies with the corneal curvature. Hence the radius of curvature of the cornea can be calculated based on image size from the anterior corneal surface.[4]
In a keratometer, the relationship between the object, image, radius of refractive surface, and distance between the surface and object is denoted as r = 2ul/ o, where r is the radius of the reflective surface, o is the object size, l is the image size, and u is the distance measured between the reflective surface and object.[5] However, the exact measurement of image size poses a problem since it is difficult to entirely still a living eye while the image is being observed. The various type of manual keratometers can be Helmholtz keratometer, Bausch and Lomb, and Javal Schiotz keratometer.[6]
There can be surgical or operating keratometers as well as automated keratometers. The activity describes various types of keratometers, parts of a keratometer, indications of keratometry, limitations, complications, patient safety, education, and clinical significance of a keratometer.[7]
Procedures
Manual Keratometers
Helmholtz Keratometer
This consists of two plates. The image is displaced half of its length by both the plates, and total changes in distance or displacement give the size of the image. This doubling of the image helps with the necessity of not moving the living eye. If there is eye movement, the image will also move together during the process. Therefore, any difficulty in the adjustment of the eye is avoided. The plates used are made of glass and are a known thickness and refractive index. These plates are placed adjacent to each other so that each plate cover half of the object of a short telescope. The prime necessity is that the cylinder axis should coincide with the plane of separation of the glass plates. The glass plates can be inclined at an angle, and the angle of incident light can be varied and measured from a point in front.[2]
Bausch and Lomb Keratometer
This keratometer is based on the principle of constant object size and image size variability.[8]
Optics and Parts
Object: This is a circular mire with two plus and two minus signs. A bulb or a lamp is used to illuminate the mires with the help of mirrors placed diagonally. When light from the mire strikes the patient cornea, it produces a diminished image behind it. The image acts as an object for the rest of the optical system.
Objective lens: helps focus the light from the image of the mire (that acts as an object) along the central axis.[9]
Diaphragm and doubling prisms: A four aperture diaphragm is placed near the objective lens. Two doubling prisms are placed beyond the diaphragm, one having a base up and another a base out prism. The prisms can move independently and parallel to the central axis of the instrument. When light passes through the left diaphragmatic aperture, it is made to deviate by the base-up prism above the main optical axis. Similarly, when the light passes through the right diaphragmatic aperture, it is made to deviate by the base out prism, thus placing the second image to the right of the central axis. When light passes through the upper and lower aperture, it doesn't pass through the prism, and an image is produced on the axis.
The total area of combined upper and lower apertures is the sum of each aperture separately, but the brightness of the image is equal. The upper and lower apertures act as Scheiner's disc, doubling the image size. Hence this helps in continuous monitoring of the correct focus. The image doubling phenomenon is a unique phenomenon for Bausch and Lomb keratometer. Double images are produced adjacent as well as 90 degrees to each other; this helps in calculating the power of the cornea in two different meridians without any rotation of the instrument. Hence, this is also labeled as one position keratometer.[1]
Eyepiece: This lens helps get a magnified view of the doubled image.
Keratometry Procedure
Instrument Calibration: The keratometer is calibrated with steel balls. A whiteboard is placed in front of the objective piece, and a black line is seen sharply focussed on it. The steel ball of the particular radius of curvature is used, the ball is placed in front of the keratometer, and the value is set on the scale. The mires are focussed on the hit and trial method by the anticlockwise and clockwise movement of the eyepiece. When the mires are focused, the instrument is said to be calibrated.[1]
Patient Adjustment: To obtain perfect keratometry, patient positioning is of utmost importance. The head should be against the headrest, and the chin should rest on the chin rest. An occluder should cover the other eye which is not being examined. This is followed by chin adjustment till the patient's pupil and the projective knob are at the same level.
Mire Focusing
After instrument calibration, alignment, and positioning of the patient, the mires are focused in the corneal center.
Corneal Curvature Assessment
The keratometer is focused on the corneal image so that there is no doubling of the central corneal image. To assess the curvature in the horizontal meridian, the plus sign of the central and left image is superimposed, taking into account the horizontal measuring control. Similarly, the central and upper images are taken into account for assessing the curvature in the vertical meridian. If there is oblique astigmatism, the two-plus signs will not be aligned. The instrument has to be then rotated till both plus signs are aligned.[10]
Javal Schiotz Keratometer
This instrument is based on variable object size and the constant image size principle.[11]
Optical and Parts
Object- This consists of two mires mounted on the arc which can be moved synchronously. The two mires combinedly form the object; hence the variable size is obtained by the movement. The first mire is stepped and has a green filter, and the other mire is rectangular and has a red filter. The mires are split in the center horizontally and are illuminated by a lamp. The first image obtained from the mires acts as an object for the rest of the optical axis.
The objective lens and doubling prisms
They form a doubled image of the new object (mires images created by the cornea). It uses a Wollaston prism which is a type of doubling prism. There is a fixed image doubling by the birefringent, also known as double refracting).[12]
Eyepiece Lens
It helps to obtain a magnified view of the doubled image.
The keratometry procedure of the Javal Schiotz keratometer is more or less similar to Bausch and Lomb keratometer.
Automated Keratometers
They are similar to manual keratometers. In this, the image reflected off the target is focused on the photodetector, which assesses the image size, and hence the radius of curvature is calculated. The mire is illuminated with the IR rays, and an infrared photodetector is used. The various advantages of automated keratometers are that they are compact, less time-consuming, and require less skill to operate.
The majority of the previous studies have found good accuracy using autokeratometry. In the market, the autokeratometers are available as isolated or autorefractometers of autokeratorefractometers. The various types of automated keratometers available are the IOL master, Pentacam, Orbscan, and corneal topographer.[13]
Surgical Keratometer
In this, the keratometers are attached to the microscope. The surgical keratometer helps in monitoring astigmatism while performing corneal and limbal procedures. However, the instrument has limited accuracy due to varied factors such as:
- It is challenging to align the visual axis of the patient with the optical axis of the keratometer.
- Anterior chamber air creates a second target reflection.
- The external compression on the globe results in a change in the corneal curvature.
- The calibrations for keratometers are for a fixed distance from the anterior corneal surface. Hence the different microscope lenses result in different working distances.[7]
Parts of Keratometer
- Headrest
- Chinrest
- Anteroposterior rotation axis scale
- Eyepiece
- Horizontal knob
- Vertical knob
- Focusing knob
- Locking knob
- Elevating knob
- Chin height knob
- Keratometer height knob
- Occluding shield
- Leveling sight
- Horizontal measuring drum
- Vertical measuring drum
- Axis scale[14]
Indications
There are various clinical indications of keratometers in routine ophthalmic practice:
- Astigmatism assessment: Keratometers help to assess the corneal astigmatic error[15]
- Contact lens fitting: Keratometers help to assess the anterior corneal surface, thus prove valuable for contact lens fitting[16]
- Corneal shape assessment: Keratometers help to assess the shape in keratoconus and kertaoglobus[17]
- Contact lens flexure assessment: Keratometers help assess the rigid gas permeable lens flexure.[18]
- Anisometropia: Keratometers help to differentiate between axial versus curvatural anisometropia[19]
- Refractive error assessment: Keratometers help to assess the refractive error in case of hazy media.[20]
- IOL power calculation: Keratometer helps to know the K reading, an essential component of the SRK formula for IOL power calculation.[21]
- Astigmatism monitoring: They help to assess the pre and post-surgical astigmatism[22]
Normal and Critical Findings
Relation Between Radius of Curvature and Power of Cornea
This is indicated by the following equation D= n-1/r, where D is the power of cornea in diameters, n is the refractive index of the cornea, and r is the radius of curvature. Ever since the invention of the keratometer by Helmholtz, the refractive index of the cornea has been 1.3375 for calibrating the instrument. Therefore the formula is read as D= 1.3375-1/r or 0.3375/r or r = 0.3375/D. Hence, the parameters of the keratometers are always labeled for the radius of curvature and dioptric power.[23]
Range of Keratometer
This ranges from 36 to 52 D (6.5 to 9.38 mm). The lower limit is 30D (5.6 mm), and the upper limit is 61 D (10.9 mm) using a lens of -1.0 D and +1.25 D, respectively, in front of the telescope's objective.[24]
Interfering Factors
Limitations
- The keratometer measurements are based on the false belief that the cornea is a sphere or has a spherocylindrical configuration, where the two principal meridians are at 90 degrees to each other. In reality, the cornea is aspheric in configuration.[25]
- The peripheral corneal surface is not considered while having refractive measurements. The keratometers take into account only the central curvature of 3 to 4 mm, which is a smaller portion and is a potential limitation.[26]
- The keratometer has limited accuracy and loses accuracy in the very flat or steep cornea.[27]
- Corneal irregularities like abrasion, erosions, epithelial defects, etc., will preclude the perfect keratometry reading due to irregular astigmatism.[28]
- Keratometers have assumed the index of refraction in radius to diopter conversion[29]
- It employs para axial optics to measure the corneal surface power.[23]
- It approximates the distance from the focal point to the distance to the image.
- Regular astigmatism due to one position astigmatism is also a limitation.
Factors Leading to Errors in Keratometry
- Faulty or improper calibration
- Excessive lacrimation
- Faulty lid position
- The patient has not fixed the instrument
- Abnormal patient position
- Accommodation error
- Corneal folds or distortion
- The abnormal focus of the corneal image[30]
Complications
- Improper functioning due to dust accumulation
- Damage to the eyepiece and prisms
- Friction between instrument parts due to reduced lubrication
- Damage to the bulb or fuse bulb
- Damage to on and off button and cable
- Improper calibration[31]
Patient Safety and Education
Keratometry is a routine investigation in day-to-day clinical practice. Every patient must be educated regarding the keratometry procedure and the importance of doing it. The patient must be educated regarding the position and how to align themselves with the instrument. The patient should also be educated on how the mires appear.
Clinical Significance
An ideal keratometer can measure the radius in various meridians from the axis of the cornea. The keratometers are designed to be rotated along a particular axis. The objects are labeled as mires. A double device has been incorporated to avoid errors and confusion due to the continuous movement of the eyes.[32]
Spherical Cornea
This is characterized by no difference in power between the two principal meridians, and the mire appears as a perfect sphere.[33]
Astigmatism
This is characterized by the difference in the power between the two principal meridians. As with the rule of astigmatism, horizontally oval mires are seen. As against the rule of astigmatism, vertically oval mires are seen. In oblique astigmatism, the principal meridian lies between 30 and 60 degrees and 120 and 150 degrees. The irregular anterior corneal surface has irregular mires and a doubling of mires.[34]
Keratoconus
Inclination and jumping of mires are seen, also called pulsating mires. Minification of mires is seen in advanced keratoconus due to high myopia. Significant astigmatism shows oval mires. In advanced keratoconus, wavy, irregular, and distorted mires are also seen.[35]
References
- 1.
- Cordero I. Verifying the calibration of a manual one-position keratometer. Community Eye Health. 2013;26(84):77. [PMC free article: PMC3936693] [PubMed: 24782588]
- 2.
- Godefrooij DA, Galvis V, Tello A. Von Helmholtz's ophthalmometer: historical review and experience with one of the last surviving original devices. Acta Ophthalmol. 2018 May;96(3):314-320. [PubMed: 28772001]
- 3.
- Hashemi H, Heydarian S, Khabazkhoob M, Yekta A, Emamian MH, Fotouhi A. Keratometry in children: Comparison between auto-refractokeratometer, rotating scheimpflug imaging, and biograph. J Optom. 2019 Apr-Jun;12(2):99-110. [PMC free article: PMC6449769] [PubMed: 30879970]
- 4.
- Martin R. Cornea and anterior eye assessment with placido-disc keratoscopy, slit scanning evaluation topography and scheimpflug imaging tomography. Indian J Ophthalmol. 2018 Mar;66(3):360-366. [PMC free article: PMC5859588] [PubMed: 29480244]
- 5.
- Miller JM. A handheld open-field infant keratometer (an american ophthalmological society thesis). Trans Am Ophthalmol Soc. 2010 Dec;108:77-95. [PMC free article: PMC3016080] [PubMed: 21212850]
- 6.
- Manning CA, Kloess PM. Comparison of portable automated keratometry and manual keratometry for IOL calculation. J Cataract Refract Surg. 1997 Oct;23(8):1213-6. [PubMed: 9368167]
- 7.
- Ventura L, De Groote JJ, Saia P, Faria e Sousa SJ. Keratometry device for surgical support. Biomed Eng Online. 2009 Dec 08;8:37. [PMC free article: PMC2795748] [PubMed: 19995421]
- 8.
- Douthwaite WA, Burek H. The Bausch and Lomb keratometer does not measure the tangential radius of curvature. Ophthalmic Physiol Opt. 1995 May;15(3):187-93. [PubMed: 7659418]
- 9.
- Szirth BC, Matsumoto E, Wright KW, Murphree AL. Attachment for the Bausch & Lomb keratometer in pediatrics. J Pediatr Ophthalmol Strabismus. 1987 Jul-Aug;24(4):186-9. [PubMed: 3668766]
- 10.
- Chang M, Kang SY, Kim HM. Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses? Korean J Ophthalmol. 2012 Feb;26(1):10-4. [PMC free article: PMC3268162] [PubMed: 22323879]
- 11.
- Dehnavi Z, Khabazkhoob M, Mirzajani A, Jabbarvand M, Yekta A, Jafarzadehpur E. Comparison of the Corneal Power Measurements with the TMS4-Topographer, Pentacam HR, IOL Master, and Javal Keratometer. Middle East Afr J Ophthalmol. 2015 Apr-Jun;22(2):233-7. [PMC free article: PMC4411623] [PubMed: 25949084]
- 12.
- Tennen DG, Keates RH, Montoya C. Comparison of three keratometry instruments. J Cataract Refract Surg. 1995 Jul;21(4):407-8. [PubMed: 8523284]
- 13.
- Leyland M, Benjamin L. Clinical assessment of a hand-held automated keratometer in cataract surgery. Eye (Lond). 1997;11 ( Pt 6):854-7. [PubMed: 9537145]
- 14.
- Karabatsas CH, Cook SD, Papaefthymiou J, Turner P, Sparrow JM. Clinical evaluation of keratometry and computerised videokeratography: intraobserver and interobserver variability on normal and astigmatic corneas. Br J Ophthalmol. 1998 Jun;82(6):637-42. [PMC free article: PMC1722645] [PubMed: 9797664]
- 15.
- Alpins N, Ong JKY, Stamatelatos G. Assessing Total Keratometry Astigmatism, Simulated Keratometry, and Total Corneal Topographic Astigmatism Against Manifest Refractive Cylinder. J Refract Surg. 2021 Mar;37(3):198-201. [PubMed: 34038302]
- 16.
- Rajabi MT, Mohajernezhad-Fard Z, Naseri SK, Jafari F, Doostdar A, Zarrinbakhsh P, Rajabi MB, Kohansal S. Rigid contact lens fitting based on keratometry readings in keratoconus patients: predicting formula. Int J Ophthalmol. 2011;4(5):525-8. [PMC free article: PMC3340718] [PubMed: 22553715]
- 17.
- Lauschke JL, Lawless M, Sutton G, Roberts TV, Hodge C. Assessment of corneal curvature using verion optical imaging system: a comparative study. Clin Exp Ophthalmol. 2016 Jul;44(5):369-76. [PubMed: 27381574]
- 18.
- Sorbara L, Fonn D, MacNeill K. Effect of rigid gas permeable lens flexure on vision. Optom Vis Sci. 1992 Dec;69(12):953-8. [PubMed: 1300519]
- 19.
- Palamar M, Degirmenci C, Biler ED, Egrilmez S, Uretmen O, Yagci A. Evaluation of the anatomic and refractive differences in hyperopic anisometropia. Int Ophthalmol. 2016 Dec;36(6):881-886. [PubMed: 26887566]
- 20.
- Moghadas Sharif N, Yazdani N, Shahkarami L, Ostadi Moghaddam H, Ehsaei A. Analysis of Age, Gender, and Refractive Error-Related Changes of the Anterior Corneal Surface Parameters Using Oculus Keratograph Topography. J Curr Ophthalmol. 2020 Jul-Sep;32(3):263-267. [PMC free article: PMC7382518] [PubMed: 32775801]
- 21.
- Lawless M, Jiang JY, Hodge C, Sutton G, Roberts TV, Barrett G. Total keratometry in intraocular lens power calculations in eyes with previous laser refractive surgery. Clin Exp Ophthalmol. 2020 Aug;48(6):749-756. [PubMed: 32279436]
- 22.
- Pontikos N, Chua S, Foster PJ, Tuft SJ, Day AC., UK Biobank Eye and Vision Consortium. Frequency and distribution of corneal astigmatism and keratometry features in adult life: Methodology and findings of the UK Biobank study. PLoS One. 2019;14(9):e0218144. [PMC free article: PMC6752876] [PubMed: 31536508]
- 23.
- Tang M, Chen A, Li Y, Huang D. Corneal power measurement with Fourier-domain optical coherence tomography. J Cataract Refract Surg. 2010 Dec;36(12):2115-22. [PMC free article: PMC3005697] [PubMed: 21111315]
- 24.
- Hashemi H, Yekta A, Shokrollahzadeh F, Aghamirsalim M, Ostadimoghaddam H, Hashemi A, Heydarian S, Khabazkhoob M. The Distribution of Keratometry in a Population Based Study. J Curr Ophthalmol. 2021 Jan-Mar;33(1):17-22. [PMC free article: PMC8102948] [PubMed: 34084952]
- 25.
- Mohammadi SF, Khorrami-Nejad M, Hamidirad M. Posterior corneal astigmatism: a review article. Clin Optom (Auckl). 2019;11:85-96. [PMC free article: PMC6697663] [PubMed: 31496856]
- 26.
- Cavas-Martínez F, De la Cruz Sánchez E, Nieto Martínez J, Fernández Cañavate FJ, Fernández-Pacheco DG. Corneal topography in keratoconus: state of the art. Eye Vis (Lond). 2016;3:5. [PMC free article: PMC4762162] [PubMed: 26904709]
- 27.
- Luft N, Siedlecki J, Reinking F, Mayer WJ, Schworm B, Kassumeh S, Priglinger SG, Dirisamer M. Impact of extreme (flat and steep) keratometry on the safety and efficacy of small incision lenticule extraction (SMILE). Sci Rep. 2021 Sep 08;11(1):17854. [PMC free article: PMC8426392] [PubMed: 34497302]
- 28.
- Roh HC, Chuck RS, Lee JK, Park CY. The effect of corneal irregularity on astigmatism measurement by automated versus ray tracing keratometry. Medicine (Baltimore). 2015 Apr;94(13):e677. [PMC free article: PMC4554023] [PubMed: 25837759]
- 29.
- Savini G, Hoffer KJ. Intraocular lens power calculation in eyes with previous corneal refractive surgery. Eye Vis (Lond). 2018;5:18. [PMC free article: PMC6053834] [PubMed: 30038922]
- 30.
- Mittelviefhaus H, Gentner C. [Errors in keratometry for intraocular lens implantation in infants]. Ophthalmologe. 2000 Mar;97(3):186-8. [PubMed: 10789175]
- 31.
- Piñero DP, Camps VJ, Caravaca-Arens E, de Fez D, Blanes-Mompó FJ. Algorithm for Correcting the Keratometric Error in the Estimation of the Corneal Power in Keratoconus Eyes after Accelerated Corneal Collagen Crosslinking. J Ophthalmol. 2017;2017:8529489. [PMC free article: PMC5672131] [PubMed: 29201459]
- 32.
- Ventura L, Riul C, Sousa SJ. Optical projection mires for measurements of the radii of curvature of the cornea in biomicroscopes. Appl Opt. 2007 Oct 10;46(29):7155-61. [PubMed: 17932523]
- 33.
- Al-Sayyari TM, Fawzy SM, Al-Saleh AA. Corneal spherical aberration and its impact on choosing an intraocular lens for cataract surgery. Saudi J Ophthalmol. 2014 Oct;28(4):274-80. [PMC free article: PMC4250498] [PubMed: 25473343]
- 34.
- Refai TA. Evaluation of the orientation of the steepest meridian of regular astigmatism among highly myopic Egyptian patients seeking non-ablative surgical correction of the refractive error. Electron Physician. 2015 Sep;7(5):1296-300. [PMC free article: PMC4590567] [PubMed: 26435831]
- 35.
- Espandar L, Meyer J. Keratoconus: overview and update on treatment. Middle East Afr J Ophthalmol. 2010 Jan;17(1):15-20. [PMC free article: PMC2880369] [PubMed: 20543932]
Disclosure: Bharat Gurnani declares no relevant financial relationships with ineligible companies.
Disclosure: Kirandeep Kaur declares no relevant financial relationships with ineligible companies.
- Corneal Topography.[StatPearls. 2024]Corneal Topography.Sridhar U, Tripathy K. StatPearls. 2024 Jan
- Vecto-keratometry: determination of anterior corneal astigmatism in manual keratometers using power vectors.[Expert Rev Med Devices. 2023]Vecto-keratometry: determination of anterior corneal astigmatism in manual keratometers using power vectors.Salvador-Roger R, Vila-Andrés R, Micó V, Esteve-Taboada JJ. Expert Rev Med Devices. 2023 Jun; 20(6):517-524. Epub 2023 Apr 28.
- Measurement of corneal curvature after corneal transplantation and radial keratotomy using standard and automated keratometry.[CLAO J. 1989]Measurement of corneal curvature after corneal transplantation and radial keratotomy using standard and automated keratometry.Binder PS. CLAO J. 1989 Jul-Sep; 15(3):201-6.