3.2. Phase distribution at each point on the optical axis in the case of exceeding the diffraction limit by simulation
The state of the phase distribution of the light from the microstructure beyond the diffraction limit at each point on the optical axis in relation to the two light sources and their distances was then verified by simulation.
Figure 5(a) illustrates the two-dimensional distribution of the electromagnetic field intensity near the optical axis for two light sources with a distance of 2 µm between two points that do not exceed the diffraction limit when the phase is both 0 rad. At 1 µm, immediately after emitted from the light source, the lights interfere as shown by the red dashed line in
Figure 5(b), which generates the phase distribution in the x-direction. It can be noted that the light is a plane wave with a flat phase depending on the beam width as a beam along the optical axis.
As shown in
Figure 5(a), this interfered zero-order diffracted light travels towards the lens with a slight spread along the optical axis. Consequently, as shown in
Figure 5(c), at 50 µm from the measurement object, the phase distribution of the zero-order diffracted light grows into a spherical wave whose phase is slightly delayed as it travels away from the optical axis because the light source itself is set as a spherical wave.
After passing through the lens at 90 µm, as shown in
Figure 5(d), the direction of convexity of the phase distribution changes so that the phase distribution at 50 µm is reversed by the lens. The phase advances at the periphery, away from the optical axis. It is shown that the light itself becomes a wavefront focused towards the focal point.
Then, at 107.8 nm in the back focus of the lens, the phase was almost flat near the optical axis, as shown in
Figure 5(e). Furthermore, on the image plane in
Figure 5(f), the phase distribution is found to be slightly spherical and convex in the direction of travel.
However, when the phases of the two sources shown in
Figure 6(a) are both 0 rad and their distance is 0.2 µm, the phase distribution does not become like the zero-order diffracted light of the Young fringe as it does when the diffraction limit of
Figure 5(a) is not exceeded. In that case, the phase is delayed in the periphery with a symmetrical phase distribution across the optical axis in
Figure 6(b), which shows the state at 1 µm immediately after the light source. It can be observed that the light wave is largely spread out.
However, in
Figure 6, Young’s fringe interference fringes cannot be observed, as shown in
Figure 5. Therefore, light from the two light sources diffuses immediately in the transverse and travelling direction near the light source. As a result, the change in the phase distribution of light in the direction of travel was small, and the phase distribution was averaged.
For this reason, the maximum and minimum values of the phase distribution in the direction of travelling y are ±0.1 rad as much smaller values in
Figure 6, compared to the values of π rad and -π rad in the case of
Figure 5. These indicate that the state of the spread of the phase distribution is clearly different when the distance between the light sources exceeds the diffraction limit and when it does not.
At the 50 μm position shown in
Figure 6(c), where the light has progressed further from
Figure 6(b), the phase is delayed at the position away from the optical axis and the range of the peripheral phase delay is expanded as a spherical wave with a broad spread. At the 90 μm position shown in
Figure 6(d) after the lens has passed through, contrary to
Figure 6(c), the phase at the periphery away from the optical axis increases and the phase distribution becomes a wavefront (phase distribution) that focuses towards the rear focal point. At the back focus of the lens, as shown in
Figure 6(e), the phase is a flat plane wave on the optical axis. At the imaging plane,
Figure 6(f), it can be noted that the phase is again a plane wave with a flatter phase. This state indicates that when lights from two light sources exceeding the diffraction limit are captured on the image plane, as shown in a previous study [
25], a phenomenon occurs that is in agreement with the results obtained when the phase is spatially flattened on the image plane, even if the phases of the two light sources are different.
Thus, if the phase is spatially flattened on the image plane, it can naturally be understood that the phase difference between the two points to be measured cannot be detected by observing the phase distribution on the image plane. However, it has been confirmed that in actual optical systems, as shown in
Figure 2, when a three-dimensional shape exists on the measurement object, the phase distribution can be detected, and the shape of the object can be reconstructed as a phase distribution, as shown in
Figure 7.
It should be noted that the experiments did not use spherical waves with phase-aligned wavefronts, as in the simulations. In other words, scattered light with wavefronts of random phase is used in the experiment. In addition, it should be noted that the phase distribution is not detected at the image formation position of
Figure 6 (a) in the simulation.
However, in the current experiment, the phase information was detected at a position that was not too close to the objective lens. Based on the experimental situation, it was also decided to detect the phase in the simulation at a position not too close to the objective lens and at a position where the phase changed significantly in the x-axis direction with respect to the direction of light travel. As a position suitable for this condition, the phase was detected at 90 µm shown in
Figure 6(d), which is considered to be able to detect the phase information on the measured object as unaffected by the surroundings as possible.
3.3. Phase analysis in the area between the lens position and the posterior focal point
The distribution of the electromagnetic intensity in
Figure 6(a) shows that when the distance between the two light sources is 0.2 µm, which is beyond the diffraction limit, the light from the light source passes through the lens while only the zeroth-order diffracted light spreads slightly radially from the light source. In other words, light from the light source spreads radially as a spherical wave immediately after it is emitted, as shown in
Figure 6(b) and (c). To obtain information on a spread of light using its intensity distribution, the light is focused again using a lens to form an image of the object, which is then observed as an image. However, in the analysis using the phase distribution, when the spreading light is refocused, the phase changes according to the length of the optical path through which the light has passed are superimposed on the phase distribution containing information concerning the shape of the original measurement object in the vicinity of the optical axis. Consequently, it may be difficult to detect the phase accurately based on the original shape of the measurement object.
It is thought that the effect of the optical path length of the spreading light may cause the original information to be hidden in the phase distribution owing to this problem in a new optical path length.
Therefore, it is essential to capture only the light near the optical axis of the zeroth-order diffracted light, which has as little spread as possible and is not focused by the lens in cases such as those shown in
Figure 6(a). In actual optical systems, the principle of measurement is based on speckle interferometry. Therefore, it is believed that the light incident on the lens is only detected near the optical axis passing through the pinhole. In other words, if only light on an optical axis parallel to the optical axis is detected as far as possible in
Figure 6(a), such that no additional phase element owing to differences in the optical path is added to the phase distribution of the measurement object, the observation of the phase distribution may be close to the experimental results using actual optical systems. However, it was assumed that the COMSOL used in this simulation would facilitate a pinhole setting by placing a wall with a PML setting just in front of the lens. However, the light reflected from the pinhole wall caused complex multiple reflections in the optical system because the PML settings cannot be sufficiently set up. In other words, the problem is that the pinhole setting cannot be always perfect. Therefore, it was decided here to perform a simpler process, instead of a pinhole-based simulation, by capturing only the light on the optical axis as far as the area indicated by the red dashed circle in
Figure 6(a) at 90 µm shown in
Figure 6(d), as the pinhole-based simulation is currently difficult to perform.
Under the conditions of this information collection, as shown in
Figure 8, two light sources (
Figure 8(a)) and three light sources (
Figure 8(b)) are assumed in this study. In this case, the distance between the light sources is set to ε µm, and the phase of each light source is set to α, β. Furthermore, using the knowledge that 2π rad corresponds to the light source wavelength of 532 nm, the position of the light sources is set approximately by changing the value of the phase. Specifically, it was decided to treat the light as approximated as coming from a source one-half of a wavelength in front (266 nm) when the phase advances by π rad. The spherical wave from the light source set up in this way passed through the lens, and the phase at a position at 90 µm (furthermore, the phase difference was converted into a length using the wavelength-phase relationship) was observed. In the optical system used in the experiments, the phase distribution of the object light was detected with high resolution based on the spatial fringe analysis method by interfering with the object light from the object and a plane wave as the reference light [
17]. Based on this processing approach, in the simulation, the wavefront corresponding to the reference light is calculated in advance as the phase distribution when both phases α and β of each light source in
Figure 8 are set to zero. By subtracting such a reference phase distribution from the phase distribution obtained by setting α and β, only the phase distribution originally intended to be obtained is calculated.
3.4. Simulation of two light sources with different phases
One of the phases α and β of the two light sources separated by ε = 0.2 µm shown in
Figure 8(a) was set to 0 rad, while the other was set to phase in the range of angles π/4, π/2, and 3π/4 in the first and second quadrants, where the phase change in the phase plane is relatively small.
As a result of using this condition, information on the simpler nature of the state, which is close to linearity, eliminating the complex nature, it is considered that the nature of the phase distribution beyond the diffraction limit, which is the purpose of this study, can be successfully verified by observing it without complicating it. As a concrete matter of consideration, it was examined whether the phase distribution of the wavefront detected at 90 µm when one of the values of α and β is set to 0 rad and the other changes to π/4, π/2, and 3π/4 rad.
Figure 9(a) shows the phase distribution when both α and β are set to 0 rad. This condition is used as the standard phase distribution for the reference beam. In the experiments, speckle interferometry was used to determine the displacement f(x+δx) - f(x) at each measurement point when the object was laterally shifted by performing a calculation between the two measurement points of the object. Based on the above concept, to determine the amount of change φαβ when one of α and β was changed by 0 and the other by π/4, π/2, and 3π/4 rad, respectively, the phase distribution φ00 when both α and β are set to 0 rad was obtained in advance. The value of φ00 was then subtracted from the value when one of them is changed by π/4, π/2, and 3π/4 rad, respectively. This processing accurately detects the phase change φαβ associated with a change in α and β.
Furthermore, the phase distribution in
Figure 9(a) is already convex downwards. This phenomenon can be confirmed by observing the electromagnetic field distribution at a distance of 90 µm of the measurement object in
Figure 6(a), where the light is focused towards the optical axis by the lens. The phase distribution is considered to be a downward convex curve because the phase is larger away from the optical axis, owing to the different optical path lengths between the light from the periphery of the lens, which is further away from the optical axis, and the light near the optical axis. This result is also observed in the phase distribution shown in
Figure 6(d). In other words, the phase information based on the shape of the measurement object cannot be properly extracted without processing to cancel the optical path length changes caused by lens focusing. Such processing has been realised in experiments using speckle interferometry.
Figure 9(b) shows the phase distribution when α = 3π/4 and β = 0. Since 2π rad corresponds to a wavelength of 532 nm, the phase distribution is converted to a length on the y-axis.
In this phase distribution as the object light, the angles of α and β are already different, so the phase distribution converted into the length of the measurement result is asymmetrical on the left and right. The distortion of the phase distribution in this oblique direction confirmed that the phases of the two light sources under examination were different. Furthermore,
Figure 9(c) shows the result for α=0 and β=3π/4. In this case, the angles of α and β are opposite to the result in
Figure 9(b), indicating that an inverse asymmetric phase distribution is obtained.
Figure 10 shows the results of replacing the phase difference with the length when the phase distribution is subtracted
Figure 9(a) with both α and β set to 0 from
Figure 9(b) and 9(c) with different angles of α and β.
The different phases of α and β on the left and right in Figs. 10(a) and 10(b), 0 and 3π/4, show that the slope of the detected phase distribution changes from 2.96 [nm/μm], shown in
Figure 10(a), where the wavefront travels in the right direction, to -3.13 [nm/μm], shown in
Figure 10(b), where the wavefront travels in the left direction. The absolute value of the slope changed slightly, but the sign also changed. The difference between the phases of α and β can be confirmed by checking the phase distribution at 90 µm.
Figure 11 shows the change in the slope of the phase distribution as shown in
Figure 10 when the values of phase α and β of the two light sources in
Figure 8 are varied as π/4, π/2, and 3π/4, which are set according to the size of the circles and the type of black and white colour, respectively. Then, the case of 0.05, 0.1, 0.2, and 0.5 μm on the horizontal axis of the distance between light sources is investigated. The phase of the two light sources changed, generating a change in the slope of the wavefront in accordance with the phase difference between the two light sources. Furthermore, the tilt of the wavefront increased with increasing distance between the sources, as shown on the horizontal axis.
In other words, the results in
Figure 11 show that the slope of the wavefront varies with changes in the phase of the light source and with the distance between the light sources, which is related to the position of the light source. In this case, according to the calculations presented in this study, because the changes were limited to a minute range, it was found that there is a proportional relationship between the intensity of the light sources and the slope of the wavefront. Furthermore, the changes in the phase of the light sources and the slope of the wavefront shown by the large●, large○ and small● are, for example, (Slope:5.09 nm/μm) for the large●, (Slope:2.10 nm/μm) for the large○ and further (Slope:0.87 nm/μm) for the small● shown in brackets for ε=0.5μm with β=0. It can be seen that the slope is approximately related to the magnitude of the phase α of the light source (π/4, π/2, 3π/4).
As shown in
Figure 8(a), it can be assumed that the phase distribution of light from the two light sources immediately after passing through the lens contains information regarding the shape of each reflection point on the object. Based on this property, the shape of the microstructure is considered to be detected in the experiment. In other words, it can be confirmed that there is a phase distribution in the zero-order diffracted light as reflected light from microstructures.
3.5. Simulation of different phases from three light sources
In the previous section, it was confirmed that the zero-order diffracted light from two light sources with different phases in close proximity beyond the diffraction limit contained information about the phases of the two light sources.
Therefore, different phases between the three light sources in close proximity beyond the diffraction limit were further investigated.
In other words, the case where the phase of the light source L1 in the middle is varied on the optical axis as shown in
Figure 8(b), with the phase of L2 and L3, which are respectively separated by ε, as 0 rad is considered. In the case of the two light sources in the previous section, when there was a difference in phase between the light sources, it was confirmed that the phase distribution was included in the zeroth-order diffracted light after passing through the lens because of the change in the direction of the wavefront travel. It was investigated whether the phase difference could be detected among the three light sources.
Figure 12(a),
Figure 12(b), and
Figure 12(c) show the phase distribution at 90 µm when the respective distance between the three light sources is set as ε = 0.2 µm when the phase of L2 and L3 is set as 0 rad and the phase of L1 is set as a negative delayed phase of -π/8, -π/4, and -π/2 rad. The phase of L1 is delayed in comparison with L2 and L3, which confirms that the phase distribution near the optical axis is delayed at 90 µm, resulting in a downward convex result.
In
Figure 12, 2π rad corresponds to a wavelength of 532 nm, and the phase difference between the line connecting L2 - L3 and L1 is defined as h
diff, which is converted to a length h. In this calculation, the results were considered using the phase distribution obtained by setting the phases of all light sources, L1, L2, and L3, to zero as a standard phase distribution.
First, phases L1, L2, and L3 were set to zero, as shown in
Figure 10, and then subtracted from the phase distribution when L1 was changed. On the other hand,
Figure 12(d), (e), and (f) are the results of the phase distribution at 90 µm, replaced by the length h, when L1 is set to be phase advanced by setting the phase α to have positive values of π/8, π/4, and π/2 rad compared to L2 and L3.
In this case, in contrast to
Figure 12(a), (b), and (c), the phase distribution in the vicinity of the optical axis advances with respect to the surroundings.
The length h is positive, yielding an upward convex result. In this case, the process was also performed by subtracting the reference phase distribution obtained by defining the phase of all light sources as zero. It can be confirmed that the zero-order diffracted light contains phase information of the measurement object with regard to the phases of the three light sources.
The detection of phase components in the zero-order diffracted light is considered to realise super-resolution technology using speckle interferometry.
Figure 13 shows the results of the phase advance and delay states in the near optical axis in the phase difference between L1 and L2, L3 shown on the horizontal axis, converted to the length h, with the parameters of the distance between the light sources ε set to 0.1, 0.2, and 0.5 μm as shown in
Figure 12. As the values obtained in the simulations were very small, the characteristics were compared by expressing the length values as a function of the phase change using logarithms. It can be found that the phase change increases as the distance ε between the light sources increases from 0.1 to 0.2 and 0.5 µm. It can also be shown that the detected length h increases as the phase difference increases from π/8, π/4 and π/2 rad. Furthermore, the case where L1, indicated by the white circle, is in a positive phase with respect to L2 and L3 (the shape of the measured object is protruding) is slightly larger than the case where the phase is lagging, indicated by the black circle (the shape of the measured object is depressed). This indicates that the protruding shape of the measured object is easier to detect. However, the results shown in
Figure 12 and
Figure 13 are the results of calculations based on simulations, which resulted in the detection of lengths below nm. In particular, in the case of ε = 0.1 µm, where the light sources are close together, the change in length h as the difference between L1 and L2 - L3 is extremely small. As a result of the calculation, the zero-order diffracted light was considered to contain phase information; however, the calculated value was very small.
It is thought that this problem should be considered by setting up a pinhole in front of the lens in the simulation to detect only zero-order diffracted light with no spread near the optical axis and by simulating scattered light with a larger number of ray vectors.
In this study, to understand the influence of parameters such as light source spacing on the results, it was decided to examine the results with three light sources using a logarithmic function, which provides a relatively good examination of the phenomenon, rather than directly comparing small arithmetic and large results.
Because it is currently difficult to set up pinholes in optical systems using COMSOL, in the future, consideration is being given to making the use of COMSOL, which is currently being used, more similar to actual optical systems by discussing with software development engineers the creation of pinholes that can prevent complex reflections and interference from occurring in the optical system.
3.6. Observation of an object with periodic structure
By investigating two- and three-light sources, it has been shown that phase information is contained in zero-order diffracted light. Therefore, a measured object with periodicity, as shown in
Figure 7, was next considered.
That is, a simulation was performed when an object consisting of 21 light sources with a period of 0.4μm. A phase of each source changes in the range of -π/2 to π/2, which are continuously and closely adjacent at 0.05μm beyond the diffraction limit, as shown by the red dashed line in
Figure 14. A phase distribution at a position of 90μm is detected.
In this case, the linearity of the measurement results based on the results shown in
Figure 9 and
Figure 12 was also considered, and a shape composed of a relatively small phase change was observed.
The results are shown in
Figure 15. In this case, the position of 0 on the x-axis corresponds to the position of the optical axis. The phase distribution at 90 μm, which is the standard information, is shown in
Figure 15(a) when the phase of all light sources is set to 0 rad, in the same way as in the 2- and 3-light source cases. In this case, the phase distribution is also converted to a length h by replacing 2π rad with a wavelength of 532 nm, and the value is shown in round brackets.
Figure 15(b) shows the phase distribution detected at 90 μm as well, with the phase set at each of the 21 points shown in
Figure 14. The result obtained by subtracting the standard phase distribution in
Figure 15(a) from that in
Figure 15(b) is shown in
Figure 15(c). In this case, the phase distribution according to the sinusoidal shape shown in
Figure 14 is shown in
Figure 15(c), although the phase change is very small. In this case, the detected sinusoidal bias component (indicated by the red line) is not necessarily 0, as shown in
Figure 14, but is 0.01 rad because the phase distribution is at 90 μm as the detection position. It is not possible to investigate precisely whether this result is presently caused by the value of the optical path length to 90 μm from the measurement object surface since the light passed through the lens.
Even for phase distributions that are more complicated than the two- and three-light sources, as shown in
Figure 15(c), the phase information is maintained within the zeroth-order diffracted light, and the shape of the original phase distribution can be retrieved after passing through the lens. In this case, the length at which a change is extracted is small. In addition, the period of the phase distribution shown in
Figure 15(c) is slightly larger, 0.76 μm, while the period of the periodic structure shown in
Figure 14 is 0.4 μm. Although the magnification was originally supposed to be 1.2x based on the positional relationship between the position of the measurement object and the imaging plane, the magnification was larger than that. The zero-order diffracted light shows a slight spread immediately after emitting from the light source, and although the beam is converging after passing through the lens, it still reaches 90 μm in a spreading state. Because the beam was slightly wider than the object to be measured, it was assumed that the phase distribution (shape distribution) was wider than 0.4μm. This phenomenon can be observed in the region around 90 μm in
Figure 6(a).
This problem is considered to be similar to the situation in the experimental results shown in
Figure 7, where a diffraction grating with a period of 278 nm was measured, and the measurement result was detected with a magnification of approximately 300 nm. This phenomenon of magnification change has often been observed in previous experiments. If this phenomenon can be corrected by setting up a pinhole in the simulation and if information can be extracted only in the area near the optical axis, it is possible that the change in magnification can be improved by collecting information at a position closer to the image formation position. In the future, it will be necessary to develop a super-resolution technology using the phase distribution based on more careful consideration by reexamining the simulation model using pinholes and scattered light.
However, to observe microstructures, situations exist where an objective must be used to magnify the image. In this case, only information near the lens optical axis needs to be extracted by pinhole, and as shown in
Figure 6, the situation where the magnification changes could be improved if the information is collected closer to the image formation position from the current 90 μm detection position. In the future, it will be necessary to investigate the situation in which measurement sensitivity and magnification change with respect to the detection position of the information in detail.
In this study, by detecting the phase distribution of light using a simulation, it was clarified that the optical explanation for the realisation of super-resolution is the conservation of the phase distribution with respect to the shape of the measurement object in the zero-order diffracted light.