SlideShare a Scribd company logo
KERATOMETRY
Dr Saurabh Kushwaha
Resident Ophthalmology
SCOPE
 Introduction
 Principle of Keratometry
 Methods of Kertaometery
 Automated Keratometer
 Clinical Applications
 Limitations
 Sources of errors
INTRODUCTION
 Keratometry is measurement of curvature of the
anterior surface of cornea across a fixed chord
length, usually 2-3 mm, which lies within the optical
spherical zone of cornea.
 Kerato : Cornea
 Metry : Measurement
 Expressed in Dioptric power
PRINCIPLE OF KERATOMETRY
 Anterior surface of cornea acts like a convex
mirror and forms 1st Purkinje image
Keratometry
Keratometry
Keratometry
PRINCIPLE OF KERATOMETRY
 The curvature of the anterior surface of cornea
determines the size of the 1st Purkinje image
 Greater the curvature of cornea, lesser is the
image size
 If the object size and distance is known and the
1st Purkinje image is determined, then the
curvature of anterior corneal surface can be
calculated
DOUBLING PRINCIPLE
DOUBLING PRINCIPLE
 Biprisms introduced into the optical system so that two
images are formed
 If eye moves – both moves
 The prism is moved until the images touch each other
such that lower edge of one image coincides with upper
edge of the other
METHODS OF KERATOMETERY
BAUSCH & LOMB KERATOMETER
ILLUMINATION SYSTEM
OBSERVATION SYSTEM
MIRES
 Circular
 Forms an image on the cornea
FOCUSING OF MIRES
 Instrument is correctly focused on corneal image
so that central image is no longer doubled
 Images are superimposed using the horizontal and
vertical measuring control
RECORDING OF THE CORNEAL
CURVATURE
RANGE OF KERATOMETER
 Range - 36.00 to 52.00 D
 Normal values - 44.00 to 45.00 D
 To increase the range - Place +1.25 D lens in
front of the aperture to extend range to 61D
 Place -1.00 D lens in front of the aperture to
extend range to 30 D
ASTIGMATISM
 Difference in power b/w 2 principal meridia
 Horizontally oval mires in WTR Astigmatism
 Vertically oval mires in ATR astigmatism
 Oblique astigmatism - principal meridia b/w 30º- 60º &
120º-150º
 Rotate the head of keratometer to get alignment
 Read out the axis from the amount of rotation
KERATOCONUS
 Inclination & jumpimg of mires on attempt to adjust
the mires - Pulsating mires
 Minification of mires in advanced cases (K >52 D) due
to increased amount of myopia
 Oval mires due to large astigmatism
 Irregular, wavy & distorted
mires in advanced keratoconus
JAVAL SCHIOTZ KERATOMETER
MIRES
Keratometry
OPTICAL SYSTEM
FOCUSING OF MIRES
Keratometry
Keratometry
KERATOMETERS
 Ziess Oberkochen
Keratometer
 Humphrey Keratometer
AUTOMATED KERATOMETERS
 Focuses reflected corneal image on to an
electronic photosensitive device, which instantly
records the size & computes the radius of
curvature
 Target mires are illuminated
with infrared light & an infrared
photodetector is used
ADVANTAGES OF AUTOMATED
KERATOMETERS
 Compact device
 Consumes less time
 Comparatively easy to operate
 No doubling device is needed
 Measures angle size in many meridians, so it
computes angle as well as power in many
meridians
 Absence of annoying glare of brightly
illuminated mires
CLINICAL APPLICATIONS OF
KERATOMETERS
 Helps in measurement of corneal astigmatic error
 Helps to estimate radius of curvature of the anterior
surface of cornea - Use in contact lens fitting
 Monitors shape of the cornea - Keratoconus
- Keratoglobus
 To monitor pre- & post-surgical astigmatism
 IOL power calculation
 Used for differential diagnosis of axial versus
curvatural anisometropia
LIMITATIONS OF KERATOMETRY
 Measurements of keratometer based on false assumption
that cornea is a symmetrical spherical or spherocylindrical
structure, with 2 principal meridia separated from each other
by 90°
 Measures refractive status of small central cornea (3-4 mm)
 Loses accuracy when measuring very flat or very steep
cornea
 Small corneal irregularities preclude use of keratometer
due to irregular astigmatism
 It cannot describe corneal asphericity
SOURCES OF ERRORS IN
KERATOMETRY
 Improper calibration
 Faulty positioning of patient
 Improper fixation by patient
 Localized corneal distortion
 Abnormal lid positioning
 Improper focusing of the corneal image
THANK YOU

More Related Content

Keratometry

  • 2. SCOPE  Introduction  Principle of Keratometry  Methods of Kertaometery  Automated Keratometer  Clinical Applications  Limitations  Sources of errors
  • 3. INTRODUCTION  Keratometry is measurement of curvature of the anterior surface of cornea across a fixed chord length, usually 2-3 mm, which lies within the optical spherical zone of cornea.  Kerato : Cornea  Metry : Measurement  Expressed in Dioptric power
  • 4. PRINCIPLE OF KERATOMETRY  Anterior surface of cornea acts like a convex mirror and forms 1st Purkinje image
  • 8. PRINCIPLE OF KERATOMETRY  The curvature of the anterior surface of cornea determines the size of the 1st Purkinje image  Greater the curvature of cornea, lesser is the image size  If the object size and distance is known and the 1st Purkinje image is determined, then the curvature of anterior corneal surface can be calculated
  • 10. DOUBLING PRINCIPLE  Biprisms introduced into the optical system so that two images are formed  If eye moves – both moves  The prism is moved until the images touch each other such that lower edge of one image coincides with upper edge of the other
  • 12. BAUSCH & LOMB KERATOMETER
  • 15. MIRES  Circular  Forms an image on the cornea
  • 16. FOCUSING OF MIRES  Instrument is correctly focused on corneal image so that central image is no longer doubled  Images are superimposed using the horizontal and vertical measuring control
  • 17. RECORDING OF THE CORNEAL CURVATURE
  • 18. RANGE OF KERATOMETER  Range - 36.00 to 52.00 D  Normal values - 44.00 to 45.00 D  To increase the range - Place +1.25 D lens in front of the aperture to extend range to 61D  Place -1.00 D lens in front of the aperture to extend range to 30 D
  • 19. ASTIGMATISM  Difference in power b/w 2 principal meridia  Horizontally oval mires in WTR Astigmatism  Vertically oval mires in ATR astigmatism  Oblique astigmatism - principal meridia b/w 30º- 60º & 120º-150º  Rotate the head of keratometer to get alignment  Read out the axis from the amount of rotation
  • 20. KERATOCONUS  Inclination & jumpimg of mires on attempt to adjust the mires - Pulsating mires  Minification of mires in advanced cases (K >52 D) due to increased amount of myopia  Oval mires due to large astigmatism  Irregular, wavy & distorted mires in advanced keratoconus
  • 22. MIRES
  • 29. AUTOMATED KERATOMETERS  Focuses reflected corneal image on to an electronic photosensitive device, which instantly records the size & computes the radius of curvature  Target mires are illuminated with infrared light & an infrared photodetector is used
  • 30. ADVANTAGES OF AUTOMATED KERATOMETERS  Compact device  Consumes less time  Comparatively easy to operate  No doubling device is needed  Measures angle size in many meridians, so it computes angle as well as power in many meridians  Absence of annoying glare of brightly illuminated mires
  • 31. CLINICAL APPLICATIONS OF KERATOMETERS  Helps in measurement of corneal astigmatic error  Helps to estimate radius of curvature of the anterior surface of cornea - Use in contact lens fitting  Monitors shape of the cornea - Keratoconus - Keratoglobus  To monitor pre- & post-surgical astigmatism  IOL power calculation  Used for differential diagnosis of axial versus curvatural anisometropia
  • 32. LIMITATIONS OF KERATOMETRY  Measurements of keratometer based on false assumption that cornea is a symmetrical spherical or spherocylindrical structure, with 2 principal meridia separated from each other by 90°  Measures refractive status of small central cornea (3-4 mm)  Loses accuracy when measuring very flat or very steep cornea  Small corneal irregularities preclude use of keratometer due to irregular astigmatism  It cannot describe corneal asphericity
  • 33. SOURCES OF ERRORS IN KERATOMETRY  Improper calibration  Faulty positioning of patient  Improper fixation by patient  Localized corneal distortion  Abnormal lid positioning  Improper focusing of the corneal image