- Main
Signal yields, energy resolution, and recombination fluctuations in liquid xenon
- Akerib, DS;
- Alsum, S;
- Araújo, HM;
- Bai, X;
- Bailey, AJ;
- Balajthy, J;
- Beltrame, P;
- Bernard, EP;
- Bernstein, A;
- Biesiadzinski, TP;
- Boulton, EM;
- Bramante, R;
- Brás, P;
- Byram, D;
- Cahn, SB;
- Carmona-Benitez, MC;
- Chan, C;
- Chiller, AA;
- Chiller, C;
- Currie, A;
- Cutter, JE;
- Davison, TJR;
- Dobi, A;
- Dobson, JEY;
- Druszkiewicz, E;
- Edwards, BN;
- Faham, CH;
- Fiorucci, S;
- Gaitskell, RJ;
- Gehman, VM;
- Ghag, C;
- Gibson, KR;
- Gilchriese, MGD;
- Hall, CR;
- Hanhardt, M;
- Haselschwardt, SJ;
- Hertel, SA;
- Hogan, DP;
- Horn, M;
- Huang, DQ;
- Ignarra, CM;
- Ihm, M;
- Jacobsen, RG;
- Ji, W;
- Kamdin, K;
- Kazkaz, K;
- Khaitan, D;
- Knoche, R;
- Larsen, NA;
- Lee, C;
- Lenardo, BG;
- Lesko, KT;
- Lindote, A;
- Lopes, MI;
- Manalaysay, A;
- Mannino, RL;
- Marzioni, MF;
- McKinsey, DN;
- Mei, D-M;
- Mock, J;
- Moongweluwan, M;
- Morad, JA;
- St. J. Murphy, A;
- Nehrkorn, C;
- Nelson, HN;
- Neves, F;
- O’Sullivan, K;
- Oliver-Mallory, KC;
- Palladino, KJ;
- Pease, EK;
- Phelps, P;
- Reichhart, L;
- Rhyne, C;
- Shaw, S;
- Shutt, TA;
- Silva, C;
- Solmaz, M;
- Solovov, VN;
- Sorensen, P;
- Stephenson, S;
- Sumner, TJ;
- Szydagis, M;
- Taylor, DJ;
- Taylor, WC;
- Tennyson, BP;
- Terman, PA;
- Tiedt, DR;
- To, WH;
- Tripathi, M;
- Tvrznikova, L;
- Uvarov, S;
- Verbus, JR;
- Webb, RC;
- White, JT;
- Whitis, TJ;
- Witherell, MS;
- Wolfs, FLH;
- Xu, J;
- Yazdani, K;
- Young, SK;
- Zhang, C
- et al.
Published Web Location
https://doi.org/10.1103/physrevd.95.012008Abstract
This work presents an analysis of monoenergetic electronic recoil peaks in the dark-matter-search and calibration data from the first underground science run of the Large Underground Xenon (LUX) detector. Liquid xenon charge and light yields for electronic recoil energies between 5.2 and 661.7 keV are measured, as well as the energy resolution for the LUX detector at those same energies. Additionally, there is an interpretation of existing measurements and descriptions of electron-ion recombination fluctuations in liquid xenon as limiting cases of a more general liquid xenon recombination fluctuation model. Measurements of the standard deviation of these fluctuations at monoenergetic electronic recoil peaks exhibit a linear dependence on the number of ions for energy deposits up to 661.7 keV, consistent with previous LUX measurements between 2 and 16 keV with H3. We highlight similarities in liquid xenon recombination for electronic and nuclear recoils with a comparison of recombination fluctuations measured with low-energy calibration data.
Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
Enter the password to open this PDF file:
-
-
-
-
-
-
-
-
-
-
-
-
-
-