Paper
13 March 2013 Evaluation of 3D-2D registration methods for registration of 3D-DSA and 2D-DSA cerebral images
Author Affiliations +
Proceedings Volume 8669, Medical Imaging 2013: Image Processing; 866931 (2013) https://doi.org/10.1117/12.2007009
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
Recent C-arm systems used for endovascular image-guided interventions enable the acquisition of three-dimensional (3D) and dynamic two-dimensional (2D+t) images in the same interventional suite. The 3D images are used to observe the vascular morphology while the 2D+t images show the current state of the intervention. By spatial alignment of 3D and 2D+t images one can facilitate the endovascular interventions, e.g. by displaying the intra-interventional tools and contrast-agent flow in the augmented 3D+t images. To achieve the spatial alignment several 3D-2D registration methods were proposed that are concerned with finding the rigid-body parameters of the 3D image. Meanwhile, the pose of the C-arm system is usually obtained through a dedicated C-arm calibration. In practice, the calibrated C-arm pose parameters are typically valid only if the imaged object is positioned in the C-arm’s isocenter. To compensate this, the 3D-2D registration should search simultaneously for the rigid-body as well as the C-arm pose parameters. For verification, we tested three 3D-2D registration methods on real, clinical 3D and 2D+t angiographic images of twenty patients, ten of which were imaged with attached fiducial markers to obtain a “gold standard” registration. The results indicate that, compared to searching solely the rigid-body parameters, by searching simultaneously for rigid-body and the C-arm pose parameters significantly improves the accuracy and success rate of 3D-2D registration methods. Among the three tested methods the intensity-based method using mutual information was the most robust, as it successfully registered all clinical datasets, and highly accurate, as the maximal fiducial registration error was less or equal than 0.34 mm.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Uroš Mitrović, Žiga Špiclin, Boštjan Likar, and Franjo Pernuš "Evaluation of 3D-2D registration methods for registration of 3D-DSA and 2D-DSA cerebral images", Proc. SPIE 8669, Medical Imaging 2013: Image Processing, 866931 (13 March 2013); https://doi.org/10.1117/12.2007009
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image registration

3D image processing

Gold

Imaging systems

Calibration

3D imaging standards

3D acquisition

Back to Top