Next Article in Journal
Modular AUV System with Integrated Real-Time Water Quality Analysis
Next Article in Special Issue
Spectral Imaging Analysis for Ultrasensitive Biomolecular Detection Using Gold-Capped Nanowire Arrays
Previous Article in Journal
An Evaluation System for the Contact Electrification of a Single Microparticle Using Microelectromechanical-Based Actuated Tweezers
Previous Article in Special Issue
Electron-Beam-Lithographed Nanostructures as Reference Materials for Label-Free Scattered-Light Biosensing of Single Filoviruses
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

A Molecularly Imprinted Polymer on a Plasmonic Plastic Optical Fiber to Detect Perfluorinated Compounds in Water

1
Department of Engineering, University of Campania “Luigi Vanvitelli”, via Roma 29, 81031 Aversa, Italy
2
Copernico S.r.l., Via Monte Hermada 75, 33100 Udine, Italy
3
IREA-CNR, Via Diocleziano 328, 80124 Napoli, Italy
*
Author to whom correspondence should be addressed.
Sensors 2018, 18(6), 1836; https://doi.org/10.3390/s18061836
Submission received: 8 May 2018 / Revised: 4 June 2018 / Accepted: 4 June 2018 / Published: 5 June 2018
(This article belongs to the Special Issue Label-free Optical Nanobiosensors)

Abstract

:
A novel Molecularly Imprinted Polymer (MIP) able to bind perfluorinated compounds, combined with a surface plasmon resonance (SPR) optical fiber platform, is presented. The new MIP receptor has been deposited on a D-shaped plastic optical fiber (POF) covered with a photoresist buffer layer and a thin gold film. The experimental results have shown that the developed SPR-POF-MIP sensor makes it possible to selectively detect the above compounds. In this work, we present the results obtained with perfluorooctanoate (PFOA) compound, and they hold true when obtained with a perfluorinated alkylated substances (PFAs) mixture sample. The sensor’s response is the same for PFOA, perfluorooctanesulfonate (PFOS) or PFA contaminants in the C4–C11 range. We have also tested a sensor based on a non-imprinted polymer (NIP) on the same SPR in a D-shaped POF platform. The limit of detection (LOD) of the developed chemical sensor was 0.13 ppb. It is similar to the one obtained by the configuration based on a specific antibody for PFOA/PFOS exploiting the same SPR-POF platform, already reported in literature. The advantage of an MIP receptor is that it presents a better stability out of the native environment, very good reproducibility, low cost and, furthermore, it can be directly deposited on the gold layer, without modifying the metal surface by functionalizing procedures.

1. Introduction

PFAs have been widely used for the last four decades in many industrial sectors and their dispersion in water has been recognized as highly dangerous for eco-systems, biodiversity and human health. The EU directive 2013/39/UE lists PFAs among the priority substances to be completely eliminated within the next 20 years, thus making this issue extremely urgent.
PFOA and PFOS are the most extensively investigated PFAs, because human exposure can occur through different pathways, although dietary intake seems to be their main route of exposure [1].
These contaminants are very persistent and refractory to different biological and chemical treatments and their presence in the environment can give rise to toxicity and bio-accumulative effects, particularly to mammalian species.
Immunotoxic effects of perfluorinated alkylated substances to cellular systems and animals are widely demonstrated [2,3], and different epidemiologic research studies have shown the potential effects of these chemical compounds on various human immune diseases.
The conventional proposed analytical methods are based on chromatographic techniques coupled with mass spectrometry [4,5,6,7,8]. Furthermore, sensors based on electrochemical and colorimetric approaches have also been described [9]. All of the mentioned methods are time-consuming, expensive and they often require a complicated pre-treatment step. In order to beat these drawbacks, it is needed to find a rapid, simple and sensitive method for the detection of perfluorinated alkylated substances.
In PFOA, PFOS or total PFAs detection, a very attractive perspective is the use of a platform based on optical fibers for fast in situ and/or remote-controlled detection. For different applications, biosensors in optical fibers allow for remote sensing and for reduced dimensions and price of the whole sensor system [10,11,12,13]. In particular, several review papers describe plasmonic optical fiber sensor platforms and their applications [14,15,16,17,18,19].
On this line of argument, we exploited a low cost surface plasmon resonance (SPR) sensor platform, based on plastic optical fibers (POFs) [20], together with a novel biomimetic polymer for the detection of PFOA/PFOS in an aqueous medium. POFs are particularly advantageous due to their easily handling and installation procedures, large diameter of the fiber (a millimetre or more), low-cost and simplicity in manufacturing [21,22,23]. In a previous work, Cennamo et al. [24] built an SPR-POF sensor based on bio-receptors obtaining an LOD of 224 ppt. In this work, a new synthetic receptor, specifically designed to recognise C4 to C12 PFAs, is used with the same SPR-POF platform reaching a better LOD (130 ppt). This result could be considered of interest when compared to the detection limit of PFAs obtained by using different approaches, as reported in Oughena et al. [25] and Trojanowicz et al. [26] or Cennamo et al. [27].
The molecular imprinting technique is a convenient tool for the preparation of molecular-recognition materials characterized by good chemical stability and selectivity. Molecular imprinted polymers are biomimetic materials imprinted with a template molecule for the purpose of retaining a memory of that specific analyte (or a specific class of molecules). MIPs exhibit many favourable aspects with respect to bio-receptors, such as an easier and faster preparation, the possibility of application outside the laboratory, for example under environmental conditions, a longer durability. Moreover, the advantage of MIPs is that they can be directly deposited on a flat gold surface by a spin coater machine without modifying the surface (functionalization and passivation), as needed for bio-receptors [24].

2. Materials and Methods

2.1. Materials

Reagents: (Vinylbenzyl)trimethylammonium chloride [CAS 26616-35-3] (VBT), 2,2-azobisiso-butyronitrile [CAS 78-67-1] (AIBN), 1H,1H,2H,2H-perfluorodecyl acrylate [CAS 27905-45-9] (PFDA) were obtained from Sigma–Aldrich (Saint Louis, MO, USA) and used without any further purification. Ethylene glycol dimethacrylate [CAS 97-90-5] (EDMA) (Sigma–Aldrich) were distilled under vacuum prior to use in order to remove stabilizers.
A certified reference material is also used to prepare the standards for dose/response curve: CRM ref n. CPA 98FE.1.N.1.5 (CPAchem Ltd., Stara Zagora, Bulgaria) a mixture of 11 components (perfluoropentanoic acid [CAS 2706-90-3], undecafluorohexanoic acid[CAS 307-24-4], perfluoroheptanoic acid [CAS 375-85-9], perfluorooctanoic acid [CAS 335-67-1], perfluoro-nonanoicacid [CAS 375-95-1], perfluorodecanoic acid [CAS 335-76-2], perfluoroundecanoic acid [CAS 2058-94-8], nonafluoro-1-butanesulfonic acid [CAS 375-73-5], perfluorooctanoate sulfonic acid [CAS 1763-23-1], heptafluorobutyric acid [CAS 375-22-4], tricosafluorodecanoic acid [CAS 375-22-4]).
All other chemicals were of analytical reagent grade. The solvent was deionised water. Stock solutions were prepared by weighing the solids and dissolving in ultrapure water (Milli-Q®, Merck KGaA, Darmstadt, Germany).

2.2. Production of MIP for PFOA and NIP

The prepolymeric mixture for MIP was prepared according to a previously optimized procedure, based on ammonium perfluorooctanoate (FPO-NH4) as the template, VBT and PFDA as the functional monomers, EDMA as the cross-linker and AIBN as the radicalic initiator. The reagents were mixed at the following molar ratio 1(Template):4(VBT):5(PFDA):50(EDMA). The mixture was uniformly dispersed by sonication (visually homogeneous milky solution). Deionised water was added to dissolve all reagents (volume ratio H2O:EDMA = 1:17.5). Finally, the AIBN was added to the solution in non-stoichiometric ratio. Also, a second monomeric solution was prepared. The composition was the same as previously described but without adding any template, in order to obtain an NIP (non-imprinted polymer).

2.3. Optical Sensor Platform

The surface plasmon resonance (SPR) sensor is based on a D–shaped POF with an optical buffer layer (Microposit S1813, MicroChem Corp., Westborough, MA, USA) between the exposed POF core and the thin gold film. This optical platform is realized by removing the cladding of POF (along half circumference), spin coating the buffer layer on the exposed core and, finally, sputtering the gold film (see Figure 1). The plasmonic sensing area is about 10 mm in length. In the visible range of interest, the buffer layer (the photoresist Microposit S1813) presents a higher refractive index than the one of the POF core. This optical buffer layer improves the performances of the SPR sensor [20]. The size of the POF is 980 μm of core (PMMA) and 10 μm of cladding (fluorinated polymer), whereas the multilayer on D-shaped POF presents a thickness of the buffer layer of about 1.5 μm and a thin gold film of 60 nm.
As shown in Figure 1, the planar gold surface can be employed for depositing the MIP receptor layer, as we will explain in the following section. In this case, the selective detection of the analyte is possible. The outline of all the production steps, from the polishing step to the MIP deposition, with the experimental setup are summarized in Figure 1.

2.4. The Experimental Equipment

The simple and low-cost experimental setup is based on a halogen lamp (HL–2000–LL, Ocean Optics, Dunedin, FL, USA), as the light source, the SPR-POF sensor and a spectrometer (FLAME-S-VIS-NIR-ES, Ocean Optics, Dunedin, FL, USA) connected to a PC. The wavelength emission range of the halogen lamp goes from 360 nm to 1700 nm, whereas the spectrometer presents a detection range from 350 nm to 1023 nm (see Figure 1).
The SPR curves, along with data values, were displayed online on the computer screen and saved with the help of the advanced software provided by Ocean Optics. The SPR transmission spectra, normalized to the reference spectrum, achieved with air as the surrounding medium, are obtained using the Matlab software (MathWorks, Natick, MA, USA). The Hill fittings of the experimental values are obtained through OriginPro software (Origin Lab. Corp., Northampton, MA, USA).
The resin block of the SPR-POF sensor is fixed on the optical table. Every time, after that the SPR curve in air (reference spectrum) is acquired, the measurements are obtained without moving the chip. If the chip sensor is moved, the reference spectrum must be acquired again.

2.5. Deposition of the MIP and NIP Layer

The MIP and the NIP layers were deposited as hereafter described. The planar sensing area (the gold surface) was washed with ethanol, then dried in a thermostatic oven at 60 °C prior to deposition of the polymer layers (MIP or NIP).
For both layers, MIP and NIP, 50 μL of the prepolymeric mixture were dropped over the sensing region (SPR surface) of the chip and spun for 80 s at 1500 rpm.
For both the polymer layers, the thermal polymerization was then carried out for 16 h at 74 °C.
The obtained polymeric film was washed and the template molecule was extracted, leaving the imprinting sites free for rebinding.
The washing and extraction procedures were characterized by two steps.
In the first step, the MIP and NIP layers were washed with 96% v/v ethanol in order to remove not-polymerized monomers residue.
In a second step, the template was extracted from MIP by washing with HCl solution (2% w/w) and 96% v/v ethanol.
The first step is conducted flushing 5 mL of ethanol on the platform and second step flushing 1.5 mL of HCl solution, 5 mL of ethanol, 1.5 mL of HCl and 5 mL of ethanol. Finally, the sensor was flushed with deionised water and dried at room temperature.

2.6. Binding Experiments

The experimental results were collected by the SPR-POF-MIP sensor and the previously illustrated measurement setup. After each addition of the sample (solution with different concentration of the analyte), we have used a standard measuring protocol based on the following three steps: first, incubation step for chemical-interaction between analytes and MIP receptor (for 10 min at room temperature); second, washing step with water (blank); third, recording step for the spectrum, when water (blank) is present as the bulk. This protocol is necessary in order to measure the shift of the resonance determined by the specific binding (analyte/receptor interaction) on the sensing surface, and not by the changes of the bulk refractive index or by non-specific binding between gold surface and analyte.
Finally, we have obtained different results exploiting a platform based on SPR-POF-NIP sensor and the same measurement set-up as above. In particular, we deposited the NIP layer on the same D-shaped POF platform. In this case, we used the same values of the PFOA concentrations and the same three steps used in the binding experimental (SPR-POF-MIP sensor): incubation step (10 min at room temperature); washing step (with water); recording step for the spectrum, when water is present as the bulk.

3. Results

3.1. PFAs Detection

Figure 2 shows the transmission spectra of the SPR-POF-MIP normalized to the reference spectrum (spectrum achieved with air as the surrounding medium), obtained by incubating solutions at increasing concentrations of PFOA in water solution (range 0–4 ppb).
In an SPR-POF platform when the refractive index at the gold–dielectric interface increases, according to SPR phenomenon theory, the resonance wavelength is shifted to the right [20].
When an MIP receptor layer is present on the gold film, the penetration of the analyte in the MIP layer produces a change (usually an increase) in the resonance wavelength due to the variation (usually an increase) of the refractive index at the interface between the MIP layer and the gold film.
As shown in Figure 2, in this case, the resonance wavelength is shifted to smaller values by increasing the concentration of PFOA in water solution. A shift like this means that, when the PFOA interacts with the MIP receptor, the refractive index value of the MIP layer decreases. This phenomenon is also present when the PFOA interacts with the antibody (bio-receptor) on the same SPR-POF platform [24].
This effect is related to the chemical composition of the perfluorinated compounds. We verified this behaviour measuring the refractive index at high concentrations of PFOA in water solutions, by an Abbe refractometer. We found that when the PFOA concentration greatly increases in the water, the refractive index of the water solution slightly decreases.
Therefore, in order to exclusively measure the shift of the resonance determined by the specific binding (analyte/MIP) on the sensing surface, and not by the changes of bulk refractive index, we used all the three previously described steps: incubation step, washing step with water, and spectrum recording step when the water (blank) is present as the bulk.
Exploiting the resonance wavelengths plotted in Figure 2 and Figure 3 reports the resonance wavelength shift, with respect to the blank (PFOA 0 ppb), versus PFOA concentration, in a semi-log scale, along with the Hill fitting to the experimental data. Each experimental value is the average of 5 subsequent measurements and the respective standard deviations (error bars), are shown as well.
Figure 4 shows the dose-response curve, with the Hill fitting to the experimental data, acquired when the PFAs compounds are present in a mix standard (a certified reference material containing 11 different PFAs (C4–C11)). As it will be shown in Discussion section, the performances obtained in the PFOA or PFAs detection are the same. The total resonance wavelength variation (∆λmax) in Figure 4 is a bit different with respect to that reported in Figure 3, because when the dimension/weight of the analyte changes the refractive index variation in the MIP layer changes.

3.2. No Binding Detection

In order to verify the non-specific binding between the sensing layer and analyte, the response of SPR-POF-NIP sensor was tested. Figure 5 shows the SPR curves at different concentrations of PFOA (0–4 ppb). When the PFOA concentration increases, the shift of the resonance wavelength is not present.

4. Discussion

4.1. Analysis of the Dose-Response Curve

The Hill fittings reported in Figure 3 and Figure 4 are obtained through OriginPro software and the parameters, obtained with the associated standard errors, are listed in Table 1.
The Hill’s equation, used in the fitting of data, is reported in the following:
Δ λ c   = λ c λ 0 =   Δ λ m a x   ·   c n ( K n   +   c n )
where c is the analyte concentration, λc is the resonance wavelength at the concentration c, λ0 is the resonance wavelength at zero concentration (blank), ∆λmax is the maximum value of ∆λc (calculated by the saturation value minus the blank value), whereas n and K are the Hill constants and they can also have a physical meaning, as it will be discussed below. Standardization curves like the one reported in Figure 3 and Figure 4 are commonly used for chemo and biosensors, and their physical meaning can be related to the fact that the absorption takes place by combination at specific sites, when the number of receptor sites available for the combination with the substrate is limited [28]. In that case, the adsorption takes place according to the Langmuir absorption isotherm, as previously reported in case of a different MIP based sensor [21]. Moreover, the parameter n in the Langmuir model is equal to 1, which has been here experimentally found.
In this section, we present a comparison between the experimental results obtained in this work and the results obtained with the same SPR D-shaped POF platform but with a bio-receptor (antibody) for PFOA [24]. From Equation (1), it is possible to notice that, if n ≈ 1 and at low concentration, i.e., at c much lower than K, the dose-response curve is linear, with sensitivity ∆λmax/K, defined as the “sensitivity at low concentration”, as shown in Equation (2):
λ c λ 0 = Δ λ c = Δ λ m a x K · c
From Table 1, for the SPR-POF-MIP sensor we obtained the sensitivity at low concentration and the LOD, for both PFOA and PFAs, in water solution. Table 2 reports the obtained “sensitivity at low concentration” and the LOD (for PFOA and PFAs detection). The parameters, obtained by the same SPR-POF platform with an antibody for PFOA [24], are also reported in Table 2 for comparison purposes. The LOD can be calculated as the ratio of three times the standard deviation of the blank and the sensitivity at low concentration (Δλmax/K) [23].
Table 2 clearly shows that the same performance obtained with an SPR-POF platform with a bio-receptor for PFAs is obtained by this SPR-POF-MIP sensor. As previously stated, the advantage of MIPs is that they can be directly deposited on the gold surface, without modifying the surface. Moreover, the MIPs are synthetical receptors presenting a number of favorable features for sensing in comparison to bio-receptors, such as a better stability out of the native environment, reproducibility and low cost.

4.2. Surface Characterization by SPR Approach

In optical sensors based on SPR in a D-shaper POF, as previously described, when the refractive index at the gold–dielectric interface increases, the resonance wavelength is shifted to the right [21,22,23]. This can be exploited to monitor the deposition process of the receptor layer (MIP receptors or Bio-receptors), since the presence of the receptor on the gold film produces an evident change in the resonance wavelength due to the variation of the refractive index at the interface between dielectric layer and the thin gold film. Figure 6 shows the resonance wavelength, when the water is present as the bulk, in the following cases: the gold surface without a receptor layer (bare surface), the gold surface with a bio-receptor for PFOA [24], the gold surface with an MIP receptor and, finally, the gold surface with an NIP receptor. The experiments were performed at room temperature and each sample was incubated 10 minutes before acquiring the signal. A shift is clearly shown in Figure 6, the refractive index of the NIP is larger than the MIP’s one, while the MIP refractive index itself is larger than the bio-receptor’s one.
Therefore, the immobilization of the bio-receptor or the deposition of the MIP/NIP layer on the sensor surface (gold film) can be directly monitored by SPR measurements exploiting the same optical platform.
In the future, we will characterize the MIP/NIP layer by SEM; meanwhile, we have shown how preliminary information about the MIP/NIP layer on the gold surface can be estimated by SPR curves (by their shape and position of the dip).

5. Conclusions

We have designed, realized and tested a novel MIP receptor for PFAs sensing by a plasmonic fiber sensor. This chemical optical sensor system is selective and able to sense very low concentrations of PFAs, with an LOD down to 0.13–0.15 ppb. The performances of the proposed system are comparable to those obtained by a specific antibody for PFOA deposited on the same optical platform, but it exhibits the classic advantages of MIP receptors: low-cost and very good reproducibility, better stability out of the native environment and the possibility of directly depositing it on the gold surface, without modifying the surface itself.

Author Contributions

Conceptualization, N.C. and G.D.; Methodology, N.C., L.Z. and G.D.; Validation, N.C., G.D., G.P., A.B., C.P., F.A. and L.Z.; Formal Analysis, N.C. and G.D.; Investigation, N.C., G.D., G.P., A.B., C.P., F.A. and L.Z.; Resources, N.C., G.D., G.P., A.B. and L.Z.; Data Curation, N.C. and G.D.; Writing-Original Draft Preparation, N.C., G.D., G.P., A.B., C.P., F.A. and L.Z.; Writing-Review & Editing, N.C., G.D. and L.Z.; Visualization, N.C., G.D., G.P., A.B. and L.Z.; Supervision, G.P., A.B. and L.Z.; Project Administration, G.P., A.B. and L.Z.; Funding Acquisition, G.P., A.B. and L.Z.

Acknowledgments

This work was funded by “POR CALABRIA FESR-FSE 2014–2020—PROMOZIONE DELLA RICERCA E DELL’INNOVAZIONE—Azione 1.1.2” project.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. DeWitt, J.C.; Peden-Adams, M.M.; Keller, J.M.; Germolec, D.R. Immuno-toxicity of perfluorinated compounds: Recent developments. Toxicol. Pathol. 2012, 40, 300–311. [Google Scholar] [CrossRef] [PubMed]
  2. Corsini, E.; Sangiovanni, E.; Avogadro, A.; Galbiati, V.; Viviani, B.; Marinovich, M.; Galli, C.L.; Dell’Agli, M.; Germolec, D.R. In vitro characterization of the immunotoxic potential of several perfluorinated compounds (PFCs). Toxicol. Appl. Pharmacol. 2012, 258, 248–255. [Google Scholar] [CrossRef] [PubMed]
  3. Corsini, E.; Avogadro, A.; Galbiati, V.; Dell’Agli, M.; Marinovich, M.; Galli, C.L.; Germolec, D.R. In vitro evaluation of the immunotoxic potential of perfluorinated compounds (PFCs). Toxicol. Appl. Pharmacol. 2011, 15, 108–116. [Google Scholar] [CrossRef] [PubMed]
  4. Saito, K.; Uemura, E.; Ishizaki, A.; Kataoka, H. Determination of perfluorooctanoic acid and perfluorooctane sulfonate by automated in-tube solid-phase microextraction coupled with liquid chromatography-mass spectrometry. Anal. Chim. Acta 2010, 658, 141–146. [Google Scholar] [CrossRef] [PubMed]
  5. Young, W.M.; South, P.; Begley, T.H.; Noonan, G.O. Determination of perfluorochemicals in fish and shellfish using liquid chromatography-tandem mass spectrometry. J. Agric. Food Chem. 2013, 61, 11166–11172. [Google Scholar] [CrossRef] [PubMed]
  6. Eriksen, K.T.; Sørensen, M.; McLaughlin, J.K.; Tjønneland, A.; Overvad, K.; Raaschou-Nielsen, O. Determinants of plasma PFOA and PFOS levels among 652 Danish men. Environ. Sci. Technol. 2011, 45, 8137–8143. [Google Scholar] [CrossRef] [PubMed]
  7. Huset, C.A.; Chiaia, A.C.; Barofsky, D.F.; Jonkers, N.; Kohler, H.P.E.; Ort, C.; Giger, W.; Field, J.A. Occurrence and mass flows of fluorochemicals in the Glatt Valley watershed, Switzerland. Environ. Sci. Technol. 2008, 42, 6369–6377. [Google Scholar] [CrossRef] [PubMed]
  8. Scott, B.F.; Moody, C.A.; Spencer, C.; Small, J.M.; Muir, D.C.G.; Mabury, S.A. Analysis for perfluorocarboxylic acids/anions in surface waters and precipitation using GC-MS and analysis of PFOA from large-volume samples. Environ. Sci. Technol. 2006, 40, 6405–6410. [Google Scholar] [CrossRef] [PubMed]
  9. Chen, L.D.; Lai, C.Z.; Granda, L.P.; Fierke, M.A.; Mandal, D.; Stein, A.; Gladysz, J.A.; Bühlmann, P. Fluorous membrane ion-selective electrodes for perfluorinated surfactants: Trace-level detection and in situ monitoring of adsorption. Anal Chem. 2013, 85, 7471–7477. [Google Scholar] [CrossRef] [PubMed]
  10. Leung, A.; Shankar, P.M.; Mutharasan, R. A review of fiber-optic biosensors. Sens. Actuators B Chem. 2007, 125, 688–703. [Google Scholar] [CrossRef]
  11. Bosch, M.E.; Sánchez, A.J.R.; Rojas, F.S.; Ojeda, C.B. Recent development in optical fiber biosensors. Sensors 2007, 7, 797–859. [Google Scholar] [CrossRef]
  12. Wang, X.D.; Wolfbeis, O.S. Fiber-Optic Chemical Sensors and Biosensors (2013–2015). Anal. Chem. 2016, 88, 203–227. [Google Scholar] [CrossRef] [PubMed]
  13. Monk, D.J.; Walt, D.R. Optical fiber-based biosensors. Anal. Bioanal. Chem. 2004, 379, 931–945. [Google Scholar] [CrossRef] [PubMed]
  14. Gupta, B.D.; Kant, R. Recent advances in surface plasmon resonance based fiber optic chemical and biosensors utilizing bulk and nanostructures. Opt. Laser Technol. 2018, 101, 144–161. [Google Scholar] [CrossRef]
  15. Anuj, K.; Sharma, R.J.; Gupta, B.D. Fiber-optic sensors based on surface Plasmon resonance: A comprehensive review. IEEE Sens. J. 2007, 7, 1118–1129. [Google Scholar]
  16. Homola, J.; Yee, S.S.; Gauglitz, G. Surface plasmon resonance sensors: Review. Sens. Actuators B Chem. 1999, 54, 3–15. [Google Scholar] [CrossRef]
  17. Caucheteur, C.; Guo, T.; Albert, J. Review of plasmonic fiber optic biochemical sensors: Improving the limit of detection. Anal. Bioanal. Chem. 2015, 407, 3883–3897. [Google Scholar] [CrossRef] [PubMed]
  18. Estevez, M.C.; Otte, M.A.; Sepulveda, B.; Lechuga, L.M. Trends and challenges of refractometric nanoplasmonic biosensors: A review. Anal. Chim. Acta 2014, 806, 55–73. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  19. Klantsataya, E.; Jia, P.; Ebendorff-Heidepriem, H.; Monro, T.M.; François, A. Plasmonic Fiber Optic Refractometric Sensors: From Conventional Architectures to Recent Design Trends. Sensors 2017, 17, 12. [Google Scholar] [CrossRef] [PubMed]
  20. Cennamo, N.; Massarotti, D.; Conte, L.; Zeni, L. Low cost sensors based on SPR in a plastic optical fiber for biosensor implementation. Sensors 2011, 11, 11752–11760. [Google Scholar] [CrossRef] [PubMed]
  21. Cennamo, N.; D’Agostino, G.; Galatus, R.; Bibbò, L.; Pesavento, M.; Zeni, L. Sensors based on surface plasmon resonance in a plastic optical fiber for the detection of trinitrotoluene. Sens. Actuators B 2013, 118, 221–226. [Google Scholar] [CrossRef]
  22. Cennamo, N.; Pesavento, M.; Lunelli, L.; Vanzetti, L.; Pederzolli, C.; Zeni, L.; Pasquardini, L. An easy way to realize SPR aptasensor: A multimode plastic optical fiber platform for cancer biomarkers detection. Talanta 2015, 140, 88–95. [Google Scholar] [CrossRef] [PubMed]
  23. Cennamo, N.; De Maria, L.; D’Agostino, G.; Zeni, L.; Pesavento, M. Monitoring of Low Levels of Furfural in Power Transformer Oil with a Sensor System Based on a POF-MIP Platform. Sensors 2015, 15, 8499–8511. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  24. Cennamo, N.; Zeni, L.; Tortora, P.; Regonesi, M.E.; Giusti, A.; Staiano, M.; D’Auria, S.; Varriale, A. A High Sensitivity Biosensor to detect the presence of perfluorinated compounds in environment. Talanta 2018, 178, 955–961. [Google Scholar] [CrossRef] [PubMed]
  25. Oughena, M.; Moliner-Martinez, Y.; Pico, Y.; Campins-Falco, P.; Barcelo, D. Analysis of 18 perfluorinated compounds in river waters: Comparison of high performance liquid chromatography-tandem mass spectrometry, ultra-high-performance liquid chromatographytandem mass spectrometry and capillary liquid chromatography-mass spectrometry. J. Chromatogr. A 2012, 1244, 88–97. [Google Scholar]
  26. Trojanowicz, M.; Koc, M. Recent developments in methods for analysis of perfluorinated persistent pollutants. Microchim. Acta 2013, 180, 957–971. [Google Scholar] [CrossRef] [PubMed]
  27. Cennamo, N.; Di Giovanni, S.; Varriale, A.; Staiano, M.; Di Pietrantonio, F.; Notargiacomo, A.; Zeni, L.; D’Auria, S. Easy to Use Plastic Optical Fiber-Based Biosensor for Detection of Butanal. PLoS ONE 2015, 19, e0116770. [Google Scholar] [CrossRef] [PubMed]
  28. Burganov, B.I.; Lobanov, A.V.; Borisov, I.A.; Reshetilov, A.N. Criterion for Hill equation validity for description of biosensor calibration curves. Anal. Chim. Acta 2001, 427, 11–19. [Google Scholar] [CrossRef]
Figure 1. Production steps for realizing an SPR sensor in a D-shaped POF with an MIP receptor and outline of the experimental setup.
Figure 1. Production steps for realizing an SPR sensor in a D-shaped POF with an MIP receptor and outline of the experimental setup.
Sensors 18 01836 g001
Figure 2. SPR spectra obtained at different concentrations of PFOA in water solution (0–4 ppb) by an SPR-POF-MIP sensor. Inset: zoom of the resonance wavelengths.
Figure 2. SPR spectra obtained at different concentrations of PFOA in water solution (0–4 ppb) by an SPR-POF-MIP sensor. Inset: zoom of the resonance wavelengths.
Sensors 18 01836 g002
Figure 3. Plasmon resonance wavelength variation (∆λ), with respect to the blank, versus the concentration of PFOA (ppb) and Hill fitting to the experimental values, in semi-log scale.
Figure 3. Plasmon resonance wavelength variation (∆λ), with respect to the blank, versus the concentration of PFOA (ppb) and Hill fitting to the experimental values, in semi-log scale.
Sensors 18 01836 g003
Figure 4. Plasmon resonance wavelength variation (∆λ), with respect to the blank, versus the concentration of PFAs (ppb) and Hill fitting to the experimental values, in semi-log scale.
Figure 4. Plasmon resonance wavelength variation (∆λ), with respect to the blank, versus the concentration of PFAs (ppb) and Hill fitting to the experimental values, in semi-log scale.
Sensors 18 01836 g004
Figure 5. SPR spectra obtained at different concentrations of PFOA in water solution (0–4 ppb) by an SPR-POF platform with an NIP layer. Inset: zoom of the resonance wavelengths.
Figure 5. SPR spectra obtained at different concentrations of PFOA in water solution (0–4 ppb) by an SPR-POF platform with an NIP layer. Inset: zoom of the resonance wavelengths.
Sensors 18 01836 g005
Figure 6. SPR spectra acquired in the presence of water on different surfaces: bare, with a bio-receptor, with an MIP receptor and with an NIP layer.
Figure 6. SPR spectra acquired in the presence of water on different surfaces: bare, with a bio-receptor, with an MIP receptor and with an NIP layer.
Sensors 18 01836 g006
Table 1. Hill parameters (SPR-POF-MIP sensor).
Table 1. Hill parameters (SPR-POF-MIP sensor).
λ0 [nm]∆λmax [nm]KnStatistics
AnalyteValueStandard ErrorValueStandard ErrorValueStandard ErrorValueStandard ErrorReduced Chi-SqrAdj. R-Square
PFOA (Figure 3)−0.1380.9413.8330.1080.1790.0601.5370.4111.0750.995
PFAs (Figure 4)−0.2770.9227.1200.2640.3890.0692.5060.70711.2380.984
Table 2. PFOA and PFAs detection in water by an SPR-POF-MIP sensor and, for comparison, PFOA detection by [24] (an SPR-POF sensor with a bio-receptor).
Table 2. PFOA and PFAs detection in water by an SPR-POF-MIP sensor and, for comparison, PFOA detection by [24] (an SPR-POF sensor with a bio-receptor).
ReceptorParametersValue
MIP ReceptorSensitivity at low c of PFOA [nm/ppb]22.14
Sensitivity at low c of PFAs [nm/ppb]18,99
LOD [ppb]
(3 × standard deviation of blank/ sensitivity at low c of PFOA)
0.13
LOD [ppb]
(3 × standard deviation of blank/ sensitivity at low c of PFAs)
0.15
Antibody [24]Sensitivity at low c of PFOA [nm/ppb]29.82
LOD [ppb]
(3 × standard deviation of blank/sensitivity at low c of PFOA)
0.24

Share and Cite

MDPI and ACS Style

Cennamo, N.; D’Agostino, G.; Porto, G.; Biasiolo, A.; Perri, C.; Arcadio, F.; Zeni, L. A Molecularly Imprinted Polymer on a Plasmonic Plastic Optical Fiber to Detect Perfluorinated Compounds in Water. Sensors 2018, 18, 1836. https://doi.org/10.3390/s18061836

AMA Style

Cennamo N, D’Agostino G, Porto G, Biasiolo A, Perri C, Arcadio F, Zeni L. A Molecularly Imprinted Polymer on a Plasmonic Plastic Optical Fiber to Detect Perfluorinated Compounds in Water. Sensors. 2018; 18(6):1836. https://doi.org/10.3390/s18061836

Chicago/Turabian Style

Cennamo, Nunzio, Girolamo D’Agostino, Gianni Porto, Adriano Biasiolo, Chiara Perri, Francesco Arcadio, and Luigi Zeni. 2018. "A Molecularly Imprinted Polymer on a Plasmonic Plastic Optical Fiber to Detect Perfluorinated Compounds in Water" Sensors 18, no. 6: 1836. https://doi.org/10.3390/s18061836

APA Style

Cennamo, N., D’Agostino, G., Porto, G., Biasiolo, A., Perri, C., Arcadio, F., & Zeni, L. (2018). A Molecularly Imprinted Polymer on a Plasmonic Plastic Optical Fiber to Detect Perfluorinated Compounds in Water. Sensors, 18(6), 1836. https://doi.org/10.3390/s18061836

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop