1. Introduction
Wireless devices have pushed forward medical science to an advanced level in which people have access to a personalized drug delivery, a remote healthcare including simple diagnostics and data-logging operations outside of hospitals, and a continuous monitoring of biophysical parameters such as blood pressure, body temperature, breathing rate, etc. Low-cost miniature technologies help to prevent sudden infant death syndrome, heart-related diseases and provide minimal invasive continuous monitoring.
Several authors studied different techniques for designing smart textiles to measure breath pattern. For example, Skrzetuska and Wojciechowski [
1] studied the ability of T-shirts equipped with a printed respiratory rate sensor to monitor the breathing pattern of two volunteers and the influence of the environmental humidity and temperature on the output of the sensor. The authors have tried several configurations of printed sensor and identified the most optimal shape and size of the sensor. The breathing of volunteers were monitored during physical activities and rest. The sensing technology were able to identify breaths but the external climate conditions were found to have an effect on the accuracy of the results.
Joyashiki and Wada [
2] proposed to monitor breathing pattern by a body-conducted sound sensor placed on the neck. The performance of the sensor was compared with two other sensors, namely air-coupled microphone and acceleration sensor. A data analysis technique based on signal processing was developed. The authors came to the conclusion that body-conducted sensor performs better in four different types of the experiments. Schatz et al. [
3] studied the application of five different types of depth sensors for breathing rate monitoring and usage of this data for the sleep apnea identification. All of the sensors output were compared with the reference sensor and two of the five sensors have been found appropriate for sleep apnea determination. Jayarathna et al. [
4] proposed to use carbon-balck elastomer sensor embedded on the textile to measure respiratory rate, cardiac cycle and body position during sleep. The authors achieved a precision of 88.9%–100% in respiratory rate monitoring. Klum et al. [
5] proposed to use multimodal patch stethoscope and monitor breathing rate based on the electrocardiogram and phonocardiogram derived approaches. Neural network has been applied for the data analysis and the setup was able to correctly identify most of the breaths. Roudjane et al. [
6] designed a respiratory rate monitoring sensor based on a spiral silver-doped hollow-core fiber antenna, a Bluetooth transceiver, an energy harvesting module and Raspberry Pi base module. The sensor was embedded into the T-shirt and the setup was able to detect breathing patterns of four volunteers.
Since sensors network devices such as Arduino boards, wireless transceivers which are based on networking protocols like Wi-Fi or Zigbee are considered bulky and inconvenient for daily usage, several authors have considered fiber optic devices for this application [
7]. Optical fibers are considered as fire-safe, immune to electromagnetic radiation and compact in terms of size [
8]. Due to a broad variety of advantages of the optical fibers, they have been gradually used for the smart textile applications as a sensing and monitoring technology [
9].
For example, Leal-Junior et al. [
10] presented smart textiles achieved by equipping the wearable with a polymer optical fiber sensor for simultaneous measurements of the breath and heart rates. A signal processing analysis was developed in order to obtain the required heart and breathing rates and filter the person’s body movements. The authors achieved accurate results: the error for the heart beat is below 4 beats per minute and the error for the respiratory rate is lower than 2 breaths per minute.
The same authors [
11] also presented the smart walker system, which consists of a 3D printed handle with an oximetry sensor and smart clothes equipped with the polymer optical fiber sensors. The smart walker system is capable of measuring oxygen saturation, gait cadence, breathing and heart rates with an accuracy of 99%. The system is proposed to be applied in the remote healthcare and robotics applications.
Gil et al. [
12] developed a real-time wireless body sensor network which simultaneously measures physical parameters such as electroencephalography to record a brain activity, photoplethysmography to record a pulse, electrocardiography to detect a heart activity and respiration rate. The system has four major units: multi-sensing module, digitalizing module, power model and monitoring-analyzing system. The network obtains the parameters from a human body, represents outputs in a frequency domain by using Fast Fourier Transform and finally sends them to a smartphone or a PC through a Bluetooth module. The authors succeeded in synchronizing four inputs with a current of 104.1 mA (STD is ± 18.3 mA) and a data rate of 16,384 bits which is 5.9 frames per second [
12]. On the other hand, it is stated in [
13,
14] that standard methods of respiration rate monitoring based on electronics cannot be used during daily activities because they require supporting tools which are bulky.
Massaroni et al. claim that smart textiles are obtaining widespread attention for the continuous monitoring of respiratory rate and sensing units such as fibers, yarns and threads with piezoresistive properties can be embedded into elastic belts [
15]. The latter one has been investigated in [
15], specifically, six piezoresistive elements have been proposed to monitor a respiratory physiology in the regions of abdominal rib cage, pulmonary rib cage and abdomen. The assessment of a performance of the system during quiet breathing and tachypnea indicates its capability to monitor chest movements in a wide range of frequency. Results show that for quiet breathing the difference of average frequencies between reference and sensing units is lower than 1% while for tachypnea the values are always lower than 4%. By definition, smart textiles have an ability to sense parameters based on the type of sensors integrated in it. In this context, significant development in integration of fiber optic sensors (FOS) into textiles has been driven by a great number of researchers. The use of FOS is reasonable due to its characteristics such as fast response, high sensitivity, chemical and biological inertness [
16]. Additionally, FOS do not generate a heat and are immune to electrical discharge in comparison with electrical devices.
Koyama et al. incorporated a hetero-core optical fiber into a textile for monitoring heartbeat and respiration of a healthy adult in a sitting position [
17]. The smart textile was sewn into the inner side of a wool cardigan in the region of left chest. The hetero-core optical fiber changes the light intensity of the transmitted light when hetero- core portion is induced to a bending. During the breathing, since the textile is enclosed between a wool cardigan and an adult’s body, the fiber experiences a load which is proportional to peak frequencies in the obtained spectrum. The radius of the curvature is 6 mm and the value of the applied force should not exceed 25 N in order to avoid fiber damage [
18].
Another optical fiber device which is suitable for a measurement of body parameters, including the blood pressure and stroke volume [
19,
20], the temperature change in thermal ablation [
21,
22], is Fiber Bragg Grating (FBG) sensor. The use of FBG for the respiratory rate detection purposes has been demonstrated in [
23,
24,
25]. The recent works about the integration of FBG into a textile have been done by European project OFSETH (optical fiber sensors embedded into technical textile for healthcare). The FBG allows to discriminate a strain by locating sensor belts in thoracic and abdominal regions in MR environment. Additionally, a single FBG embedded in a PVC laminate shows a sensitivity of 0.8 pm/
[
26,
27]. The sensor measures cardiac frequency and breathing rate by showing strain peaks.
The aim of this paper is to contribute to the research area related to a wearable textile technology for a real-time continuous monitoring of the biophysical parameters in order to achieve a personalized health control. This work makes use of of two arrays of five FBG sensors embedded into elastic belts, located on the chest and on the abdomen, for the respiratory rate monitoring. Each array is composed by 5 FBGs for a total of 10 points of measurement. Specifically, the proposed design aims to convert breathing movement, detected in chest and abdomen regions of volunteers, to strain values. Moreover, the good number of redundant sensing points, located in different positions, is used to better reconstruct the breathing pattern also when the breath reading is disturbed by different body movements. A specific algorithm, using a diversity technique, has been purposely developed to exploit the information coming from the 10 sensing points. By using a mobile app VibSensor, which is sensitive to vibrations, as a reference unit, the influence of volunteers’ position on the breathing rate is identified and compared with the readings coming from our system.
2. Methodology
The aim of the experiments is to study the feasibility of the FBGs arrays for breathing pattern monitoring application. The breathing pattern is measured at the two locations of the body (abdomen and chest) by two arrays of 5 FBGs.This allows to apply a diversity technique, which is used in communication systems to achieve better accuracy of detection by combining the outputs from several different sensing points [
28]. The breathing pattern has been measured in four different positions of the volunteers (sitting, lying, staying or running). Two volunteers have participated in the initial experiments: 23-years old woman, height 165 cm, weight 52 kg and 24-years old man, height 171 cm, weight 72 kg. The volunteers wore T-shirts with two specially designed belts located on the abdomen and chest regions. They have been asked to breath for 23 seconds in different positions: staying, sitting, lying and running. The results of the the strain change detected by FBGs have been compared with the output of a reference sensor, which is a mobile application for acceleration and vibration measurement. The mobile phone with the application has been attached to the upper belt.
The experimental setup, which is illustrated in
Figure 1, consists of the (1) I-MON interrogator connected to (2) PC with evaluation software, (3) T-shirt with two belts each equipped with an array of 5 FBGs, (4) mobile phone with VibSensor application.
2.1. Interrogation System
The data acquisition was performed by I-MON 512 USB (Ibsen Photonics, Sweden) interrogator, shown in
Figure 1. The schematics of the interrogator is illustrated in
Figure 2 illustrates its schematics. The interrogator consists of a superluminescent LED (SLED, Exalos EXS, 1520–1600 nm, 10 mW), a control board, which is responsible for stabilizing of driving current and operative temperature, a power supply, a spectrometer and a 50/50 coupler which delivers a light to the optical fiber with FBG sensors. I-MON 512 USB has an evaluation software which displays and saves the raw data. The wavelength fits resolution of the system is less than 0.5 pm and the measurement frequency is 3000 Hz [
29]. The software provides the wavelength spectrum as an output. MATLAB script has been developed in order to convert the wavelength spectrum to the strain change pattern. The size of the interrogator used in this experiment is 37 cm × 24 cm × 19 cm. It is our future goal to reduce the interrogator size in order to make it portable. The preliminary experiments show encouraging results.
2.2. Fiber Bragg Grating Sensors
In the experiments, two arrays of 5 FBGs sensors manufactured by FBGS International have been used. Each FBG sensor has a length of 5 mm and the center to center distance between two neighboring FBGs is 10 mm. The wavelengths of the first FBGs array are in the range from 1540 nm to 1548 nm and the space between the peaks of the neighboring FBGs is 2 nm. The FBGs of the second array are located in the range from 1570 nm to 1578 nm with the same spacing of 2 nm between the two consequent peaks.
When a spectrum of light propagates through the fiber a specific wavelength, called Bragg wavelength, is reflected whereas the remaining part of the light is transmitted [
30]. When an external temperature or strain are induced, FBGs react accordingly by causing a proportional shift of the reflected Bragg wavelengths. The Bragg wavelength is represented by the following equation [
31]:
where
is the periodicity of the gratings and
is the effective refractive index of the fundamental mode.
The sensitivity coefficients for temperature and strain of the FBGs, used in this work, have been experimentally characterized, the found values are:
= 10.2 pm/
C for temperature, and
= 1.03 pm/
for strain. The characterization curves are shown in
Figure 3. Since the FBGs arrays have been manufactured with a draw tower inscribing the FBBGs in a standard SMF-28 fiber, the characteristics of each FBGs are substantially identical, in terms of thermal and strain properties. The experimental found coefficients matched the values declared by the manufacturer.
The temperature and strain coefficients are used to express the wavelength shift, according to the following equation [
32]:
where
is the Bragg wavelength when an optical fiber in a reference position,
is the strain change and
is the temperature shift.
The fiber with five FBGs has been bent in order to achieve strain change.
Figure 4 shows change of the spectrum of the FBGs array due to this applied strain change (from strain 1 to strain 2). This resulted in the wavelength shifts detected by each FBGs differently. The first left FBG has experienced compression, so its wavelength shift is negative. The other four FBGs have been under tension and detected the positive, but different strains. According to
Figure 4, the second FBG from the left has experienced a small strain change although the fiber has been bent. This shows that some of the points on the fiber can be stretched more than others at the same moment. Therefore, the multi-point sensing is used in the experiments in order to cover more area and increase the accuracy of the breathing detection. The information from all FBGs can be analysed collectively in order to obtain the correct pattern.
2.3. Belt Design and Fibers Fixation
As can be seen in
Figure 1, there are two belts embedded into the T-shirt, one on the chest region and another one on the abdomen. The belts consist of two parts: brown fabric is inelastic, and the black fabric is very elastic. The belts need to fit around the chest and the abdomen. The FBGs are placed on the elastic part because when the person breathes, the elastic part stretches and this elongation is detected by FBGs. As a result, by this design the FBGs are able to monitor the movements of the chest and the abdomen.
Figure 5 shows the fixation of the array of 5 FBGs on the elastic part of the material. As it can be seen, the fiber is fixed at four points and the FBGs are not fixed in order to have an opportunity to move. The order of FBGs start from the left to the right: from FBG 1 to FBG 5 for the chest region and from FBG 6 to FBG 10 for the abdomen region.
2.4. Reference Sensors
To check the validity of the optical fiber sensors for the breathing pattern monitoring a reliable reference sensor is required. One of the options is to use available mobile applications which measure acceleration or vibrations. The accuracy of 12 mobile applications for vibration measurements have been studied by Cahil et al., who compared their performance with the tri-axial accelerometer [
33]. According to this study, the VibSensor application has shown a good precision, so it is selected as a reference in the experiments. The application is available on the App Store and provides the graphs of the acceleration change over time in three directions (x—left and right, y—up and down, z—back and forth). Mobile phone with the application has been positioned on the chest as shown in
Figure 1. Although, the application does not directly measure the breathing pattern, the expansion of the chest causes movement of the mobile phone, at least, in one direction. Therefore, the acceleration pattern of the mobile phone can provide the information about breathing pattern.
2.5. Algorithm for the Breathing Pattern Reconstruction
As a result of the experiments, I-MON interrogator has registered the wavelength per time of each of 10 FBGs. The algorithm, that reconstructs the breathing pattern from the wavelength pattern obtained from I-MON, consists of the following steps:
(1) Convert wavelength pattern per time to the strain change by using Equation (
2). Since temperature in this experiment is constant,
is 0. Then, the equation can be rearranged to find strain:
Figure 6 shows the strain change per time detected by each FBG.
(2) The strain change detected by FBGs is a result of not only breathing, but all the movements of the chest and abdomen including heart beat and motions of a person. Therefore, the strain has to be filtered in order to obtain the components responsible for breathing only. This has been done by converting the strain pattern to frequency domain by Fast Fourier Transform. People’s normal breathing rate is about 20 breaths per minute [
34] (0.33 Hz), so everything higher than this frequency has been filtered. Then the signal has been converted back to time domain and the strain patterns shown on
Figure 7 have been obtained.
(3) According to
Figure 7, FBGs 1, 2, 4 and 5 on the chest region have similar patterns and have detected 6 peaks: on the 3rd, 7th, 10th, 14th, 17th and 21st seconds. Third FBG also has similiar pattern but with opposite sign because while other FBGs experience tension, this FBG feels tension. FBGs 6, 7, 9 and 10 on the abdomen region have also detected 6 peaks correctly, but FBG 8 has unclear pattern. It can be seen, that using this information eight FBGs out of ten can be used for reconstruction, third FBG on the chest needs to be reflected to satisfy the others pattern. One FBG out of ten in this case has given inaccurate result being unable to correctly identify the breathing pattern.
The algorithm is aimed at finding the most common pattern, reflecting opposite strains and neglecting discrepant patterns. This is achieved by firstly identifying positive and negative peaks of all FBGs. If at least 6 FBGs out of ten detected either positive or negative peak at approximately same time, this peak is considered as correct. The strain pattern of FBGs with correct negative peaks are multiplied by −1. FBGs which have less than 30% correctly detected peaks are assumed to produce wrong pattern and is neglected by multiplying its strain change by 0.
Figure 8 illustrates the strain patterns of all FBGs after the implementation of the algorithm. It can be seen that the third FBG on the chest region has been reflected, so now its pattern corresponds to the general trend. The eighth FBG on the abdomen region has been multiplied by 0 in order not to be considered in the further analysis.
(4) The mean of all ten patterns was calculated and the single breathing pattern shown on
Figure 9 (left) has been achieved. The algorithm helped to identify 6 breaths at the following seconds: 0–4, 4–8, 8–12, 2–16, 16–20 and 20–23.
Figure 9 (right) shows the plot obtained from the reference sensor, which has also detected 6 breaths during 23 s.